In this study, we aim to clarify the dosimetric characteristics of a real time variable shape rubber containing tungsten (STR) as a thin bolus in 6-MV photon radiotherapy. The percentage depth doses (PDDs) and lateral dose profiles (irradiation field = 10 x 10 cm(2)) in the water-equivalent phantom were measured and compared between no bolus, a commercial 5-mm gel bolus, and 0.5-, 1-, 2-, and 3-mm STR boluses. The characteristics of the PDDs were evaluated according to relative doses at 1 mm depth (D-1mm) and depth of maximum dose (d(max)). To determine the distance of the shift caused by the STR bolus, the PDD value at a depth of 100 mm without a bolus was obtained. For each STR thickness, the difference between the depth corresponding to this PDD value and 100 mm was calculated. The penumbra size and width of the 50% dose were evaluated using lateral dose profiles. The D-1mm with no bolus, 5-mm gel bolus, and 0.5-, 1-, 2-, and 3-mm STR boluses were 47.6%, 91.5%, 78.2%, 86.6%, 89.3%, and 89.4%, respectively, and the respective d(max) values were 15, 10, 13, 12, 11, and 10 mm. The shifting distance of the 0.5-, 1-, 2-, and 3-mm STR boluses were 2.7, 4.4, 4.8, and 4.9 mm, respectively. There were no differences for those in lateral dose profiles. The 1-mm-thick STR thin bolus shifted the depth dose profile by 4.4 mm and could be used as a customized bolus for photon radiotherapy.
机译:在这项研究中,我们的目标是阐明剂量测定的实时变量的形状特征含钨橡胶(STR)作为薄丸在6-MV光子放射治疗。剂量(广泛性成长障碍)和横向剂量概要文件(辐照字段= 10 * 10厘米(2))水当量测量和幽灵之间没有丸相比,商业5毫米凝胶丸,0.5 - 1,2 - 3毫米STR丸。广泛性成长障碍的特点根据相对剂量在1毫米深度(D-1mm)和深度的最大剂量(d (max))。STR的距离变化引起的丸,PDD值100毫米的深度一丸。相对应的深度之间的区别这个PDD值和100毫米计算。半影50%剂量的大小和宽度评估使用横向剂量概要文件。没有丸,5毫米凝胶丸,和0.5 - 1 - 2 -和3毫米STR丸分别为47.6%、91.5%、78.2%,86.6%、89.3%和89.4%,分别各自的d (max)值是15日,10日,13日,12日,11日,和10毫米。2 -, 3毫米STR丸分别为2.7,4.4,4.8,分别为4.9毫米。对于那些在横向剂量概要文件。1-mm-thick STR薄丸转移深度剂量配置文件由4.4毫米和可以作为定制丸对光子放射治疗。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号