首页> 外文期刊>Physical and Engineering Sciences in Medicine >Patient?specific quality assurance on a Varian Halcyon linear accelerator using a PTW Octavius 4D device
【24h】

Patient?specific quality assurance on a Varian Halcyon linear accelerator using a PTW Octavius 4D device

机译:病人吗?宁静的直线加速器使用PTW屋大维4 d装置

获取原文
获取原文并翻译 | 示例
           

摘要

This study investigates and validates the use of the Octavius 4D system for patient specific quality assurance on Halcyon, which is capable of rotating at 4 revolutions per minute (RPM). A commercially available PTW Octavius 4D system was used for this study which had a maximum rotation speed of 3 RPM. Initial validation included testing the accuracy of the inclinometer, percent depth doses (PDD), output factors, and dose profiles for selected static square fields. The same static fields were also subject to a gamma comparison with the TPS. This was followed by an evaluation of twelve clinical treatment plans and seven non-clinical plans with varying gantry rotation speeds. All testing was completed using detector array measurement times of 200 ms and 100 ms. Inclinometer accuracy was within 0.3 degrees of actual gantry angle. Output factors varied less than 0.6%, PDD differences were no greater than 1.4%, and dose profile differences were less than 2.2%. Gamma pass rates for the static fields were 96.7% (2%/2mm) and 99.7% (3%/3mm). A prototype control unit, which had a maximum rotation speed of 4 RPM was also used to test the clinical and non-clinical plans. For the clinical plans, the mean gamma pass rates (2%/2mm) were 86.1% and 88.1% for the commercial unit and prototype unit respectively. Results using a measurement time of 200 ms were superior to those using 100 ms. For Halcyon deliveries greater than 3 RPM, worst case gamma results for the commercial unit were 28.6% compared to 98.5% using the prototype unit. Accurate patient specific quality assurance results can be obtained using the Octavius 4D system with a Halcyon linac, provided that the system measurement time is kept at 200 ms and the rotation speed of Halycon does not exceed 3 RPM. For higher RPM deliveries, an Octavius 4D unit with 4 RPM rotation capability is recommended.
机译:本研究探讨并验证的使用病人具体的屋大维4 d系统在宁静的质量保证,能够每分钟旋转4转(RPM)。商用PTW屋大维4 d系统用于本研究最大旋转3转的速度。测斜仪的测试精度,百分比深度剂量(PDD),输出因素,和剂量选择静态广场字段的配置文件。相同的静态字段也受到γTPS的比较。计划和评估12个临床治疗7非临床计划不同龙门旋转速度。200 ms和探测器阵列测量100 ms。度的实际龙门角。多种多样的不到0.6%,PDD没有差异大于1.4%,和剂量的差异是不到2.2%。静态字段分别为96.7%(2% / 2毫米)和99.7%(3% / 3毫米)。最大转速4 RPM也使用测试的临床和非临床计划。临床计划,意味着伽马通过率(2% / 2毫米)商业的分别为86.1%和88.1%单元测试和原型单位分别。使用一个测量时间200毫秒的优越那些使用100 ms。大于3 RPM,坏的情况下伽马的结果商业单位是28.6%到98.5%使用原型单位。具体的质量保证的结果获得使用屋大维4 d系统宁静的直线加速器,系统提供保持在200 ms,测量时间旋转的速度Halycon不超过3 RPM。更高的RPM交付,屋大维4 d单元建议4 RPM旋转功能。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号