A peritoneal dialysis catheter salvage algorithm was developed and performed for 40 patients with documented catheter malfunction (obstruction and/or malposition) referred to the interventional radiology suite. This procedure utilized a metallic stiffener for repositioning and rotating dual guide wires for recanalization. A retrospective analysis of 35 cases of fluoroscopic manipulation showed that in 83% of the cases, the catheters were successfully repositioned and/or recanalized, and in 59%, they remained patent at 30 days. No major adverse events occurred. The results suggest that this algorithm is a safe and effective approach to salvage malfunctioning peritoneal dialysis catheters and that a trial of fluoroscopic salvage can be considered prior to surgical intervention.
机译:我们开发了一种腹膜透析导管抢救算法,并对40名介入放射学套件中记录的导管故障(阻塞和/或位置不当)患者进行了抢救。该程序利用金属加强件重新定位和旋转双导丝进行再通。对35例透视操作病例的回顾性分析显示,83%的病例导管成功复位和/或再通,59%的病例在30天时仍保持未闭状态。未发生重大不良事件。结果表明,该算法是抢救故障腹膜透析导管的一种安全有效的方法,在手术干预之前可以考虑进行透视抢救试验。