During transcatheter aortic valve implantations (TAVI) and other percutaneous structural procedures, some patients may need close anesthesiological care, thus exposing the anaesthesiologist to x-rays. This work aims to investigate the radiation dose received by anaesthesiologists during these specific procedures in order to improve their radiological protection. Occupational radiation doses were measured prospectively during percutaneous structural procedures in several health professionals using electronic dosimeters worn over the apron at chest level. A sample of 49 procedures were recorded, where the anaesthesiologists' average dose per procedure resulted 13 times higher than that of interventional cardiologists. The average dose per procedure received over the protection apron during TAVIs by the anaesthesiologist was 0.13 mSv, with a maximum value of 0.69 mSv. Taking these figures as a conservative estimation of the eye lens dose, an anaesthesiologist could participate in around 150 procedures before reaching the regulatory annual dose limit for the lens of the eye in Europe (20 mSv). In those clinical procedures where patients need close anesthesiological care, the anaesthesiologists might receive high radiation doses increasing the risk for cataracts and the risk of stochastic radiation effects. The proper use of occupational dosimeters will help identify these situations. It is recommended to use a mobile shielding barrier to reduce radiation exposure to acceptable levels in these situations.
机译:在经导管主动脉瓣植入术(TAVI)和其他经皮结构手术期间,一些患者可能需要密切的麻醉护理,从而使麻醉师暴露在x射线下。这项工作的目的是调查麻醉师在这些特定程序中接受的辐射剂量,以改善他们的辐射防护。在几名卫生专业人员的经皮结构手术中,使用佩戴在胸前围裙上的电子剂量计对职业辐射剂量进行了前瞻性测量。记录了49次手术的样本,其中麻醉师每次手术的平均剂量比介入心脏病医生的平均剂量高13倍。在TAVIs期间,麻醉师在防护围裙上接受的每次手术的平均剂量为0.13 mSv,最大值为0.69 mSv。将这些数字作为对眼睛晶状体剂量的保守估计,麻醉师可以在达到欧洲眼睛晶状体的监管年度剂量限值(20 mSv)之前参与大约150个程序。在那些患者需要密切麻醉护理的临床手术中,麻醉师可能会接受高剂量的辐射,从而增加白内障的风险和随机辐射效应的风险。正确使用职业剂量计将有助于识别这些情况。在这些情况下,建议使用移动屏蔽屏障将辐射暴露降低到可接受的水平。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号