首页> 外文期刊>The journal of maternal-fetal & neonatal medicine >A randomized trial comparing the short binasal prong to the RAM cannula for noninvasive ventilation support of preterm infants with respiratory distress syndrome
【24h】

A randomized trial comparing the short binasal prong to the RAM cannula for noninvasive ventilation support of preterm infants with respiratory distress syndrome

机译:一种随机试验,将短二合一叉对呼吸窘迫综合征的早产儿的非侵入式通风支撑的RAM插管进行比较

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: In this study, we compared the efficacy of the RAM cannula and the short binasal prong (SBP) as noninvasive ventilation (NIV) interfaces in preterm infants with respiratory distress syndrome (RDS). Materials and Methods: Premature infants with RDS who required NIV were randomized in the RAM cannula and SBP groups within the first half hour. The groups were compared in terms of their need for invasive ventilation, their surfactant use, and their morbidities. Results: We assessed 126 patients (62 SBPs and 64 RAM cannulas). Clinical and demographic features of the groups were similar. Within the first 72 hours, the RAM cannula group showed a higher need for invasive ventilation (32.8 and 9.6%, p = .002, respectively), surfactant (42.1 and 19.3%, p = .007, respectively), and rate of pulmonary hemorrhage was also higher. There were no differences between the groups in terms of other morbidities. Conclusions: In preterm infants with RDS, RAM cannula use as an interface for NIV results in increased invasive ventilation and surfactant use.
机译:目的:在这项研究中,我们比较了RAM套管和短双耳尖(SBP)作为无创通气(NIV)接口在患有呼吸窘迫综合征(RDS)的早产儿中的效果。材料和方法:需要NIV的RDS早产儿在前半小时内随机分为RAM套管组和SBP组。两组患者在有创通气需求、表面活性剂使用和患病率方面进行了比较。结果:我们评估了126名患者(62个SBP和64个RAM套管)。两组患者的临床和人口学特征相似。在最初72小时内,RAM套管组对有创通气的需求更高(分别为32.8%和9.6%,p=0.002),表面活性剂(分别为42.1%和19.3%,p=0.007),肺出血的发生率也更高。在其他疾病方面,两组之间没有差异。结论:在患有RDS的早产儿中,使用RAM套管作为NIV接口会增加有创通气和表面活性物质的使用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号