Flow irregularities from syringe infusion pumps caused by syringe stiction

机译:从注射器静态引起的注射器输液泵流量不规则

获取原文
获取原文并翻译 | 示例

摘要

Objective The current study aimed to evaluate the extent of the slide-stick phenomenon in differently designed infusion syringes at various infusion rates and filling positions. Methods Fluid delivery from three 50-mL infusion syringe brands (BD; Codan; Fresenius) was investigated using a flow sensor at flow rates of 0.5, 1.0, or 5.0 mL h(-1), with the syringes filled with either 10, 30, or 50 mL of distilled water. Two identical models (A/B) of the same infusion pump model were used. The effect of flow rate variations on the plasma concentration of a continuous epinephrine infusion in a 3 kg neonate receiving a continuous infusion of 0.1 mu g kg min(-1)epinephrine was studied using a pharmacokinetic simulation model. Results Considerable variations in calculated plasma epinephrine concentration were detected between flow rates of 5 and 0.5 or 1 mL h(-1)for all syringe types and filling volumes. The median deviation of plasma concentration for the 5 mL h(-1)flow rate varied depending on assembly from 1.3% (Codan) to 1.8% (Fresenius). This was more pronounced for lower flow rates, where at 1 mL h(-1)the deviation varied from 3.3% (BD) to 4.8% (Fresenius) and at 0.5 mL h(-1)from 4.9% (BD) to 5.4% (Fresenius). Differences between filling volumes (within syringe type and flow rate) did not appear to have relevant influence on variations in calculated plasma epinephrine concentration. Conclusion Infusion set rate rather than syringe brand or filling volume was a major predictor for syringe stiction-related amount of variation in the calculated plasma epinephrine concentration.
机译:目的本研究旨在评估不同设计的输液注射器在不同的输液速率和灌装位置下的滑棒现象程度。方法使用流速为0.5、1.0或5.0ml h(-1)的流量传感器,对三种50ml输液注射器(BD、Codan、Fresenius)的液体输送进行研究,注射器中填充10、30或50ml蒸馏水。使用同一输液泵型号的两个相同型号(A/B)。使用药代动力学模拟模型研究了连续输注0.1μg kg min(-1)肾上腺素的3 kg新生儿中流速变化对连续输注肾上腺素血浆浓度的影响。结果在5和0.5或1ml h(-1)的流速之间,所有注射器类型和填充量的计算血浆肾上腺素浓度均存在显著差异。5 mL h(-1)流速的血浆浓度的中位偏差因装配而异,从1.3%(Codan)到1.8%(Fresenius)。这在流速较低时更为明显,在流速为1ml h(-1)时,偏差从3.3%(BD)到4.8%(Fresenius)不等,在流速为0.5ml h(-1)时,偏差从4.9%(BD)到5.4%(Fresenius)。充盈量(注射器类型和流速)之间的差异似乎对计算的血浆肾上腺素浓度的变化没有相关影响。结论输液器速率而非注射器品牌或充盈量是计算血浆肾上腺素浓度中注射器粘滞相关变化量的主要预测因子。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号