Accuracy assessment of bolus and basal rate delivery of different insulin pump systems used in insulin pump therapy of children and adolescents

机译:胰岛素泵治疗儿童和青少年胰岛素泵治疗的提高胰蛋白酶泵系统的准确评估

获取原文
获取原文并翻译 | 示例
           

摘要

Abstract Background Continuous subcutaneous insulin infusion (CSII) is commonly used in patients with diabetes. Accurate and reliable delivery by insulin pumps is essential for a safe and effective therapy, particularly when using small doses. In this study, accuracy of bolus and basal rate delivery of various available insulin pumps was evaluated. Methods In total, 13 insulin pump systems were tested: eight durable pumps with different infusion sets and one patch pump. Based on IEC 60601‐2‐24, insulin delivery was measured by recording weight gain of a beaker into which insulin was infused by the pumps. Bolus accuracy was determined by individually weighing 25 consecutive 0.1 or 1.0 U boluses and basal rate accuracy was determined during basal rate delivery of 0.1 or 1.0 U/h for 72?hours. For analyses, basal rate delivery was divided into 1‐hour windows and deviation from target was calculated. Results Regarding different systems, average 0.1?U bolus delivery was ?2% to +9% of the intended volume with 53% to 96% of boluses within ±15% of target. During 0.1?U/h basal rate delivery, most pumps showed an initial over‐delivery for the first few hours. Three systems reached a total basal rate error 5%; others showed up to +24%. In general, delivery was more accurate when using larger doses. Conclusions Considerable differences in insulin delivery accuracy were observed between the tested pumps. In general, when using very low doses, accuracy of insulin delivery is limited in most insulin pumps. This should be considered for CSII therapy in children.
机译:摘要背景持续皮下胰岛素输注(CSII)常用于糖尿病患者。胰岛素泵准确可靠的输送对于安全有效的治疗至关重要,尤其是在使用小剂量胰岛素时。在这项研究中,评估了各种可用胰岛素泵的推注和基础速率输送的准确性。方法共测试了13个胰岛素泵系统:8个不同输液器的耐用泵和一个贴片泵。根据IEC 60601-2-24,通过记录泵向其中注入胰岛素的烧杯的增重来测量胰岛素输送量。通过分别称量25个连续的0.1或1.0 U药丸来确定药丸的准确度,并在72小时内以0.1或1.0 U/h的基本速率给药期间确定基本速率的准确度?小时。为了进行分析,将基本速率交付划分为1小时窗口,并计算与目标的偏差。不同系统的结果平均为0.1?你送的是什么?预期体积的2%至+9%,53%至96%的丸量在目标的±15%范围内。在0.1?U/h基本速率输送,大多数泵在最初的几个小时内表现出初始过量输送。三个系统达到了总的基本速率误差;5%; 其他人则高达+24%。总的来说,使用大剂量时,分娩更准确。结论两种胰岛素泵在胰岛素输送准确性方面存在显著差异。一般来说,当使用极低剂量时,大多数胰岛素泵中胰岛素输送的准确性受到限制。在儿童CSII治疗中应考虑这一点。

著录项

  • 来源
    《Pediatric diabetes.》 |2020年第4期|共8页
  • 作者单位

    Diabetes Clinic for Children and AdolescentsMuenster Germany;

    Institut für Diabetes‐TechnologieForschungs‐und Entwicklungsgesellschaft mbH an der Universit?t;

    Institut für Diabetes‐TechnologieForschungs‐und Entwicklungsgesellschaft mbH an der Universit?t;

    Institut für Diabetes‐TechnologieForschungs‐und Entwicklungsgesellschaft mbH an der Universit?t;

    Institut für Diabetes‐TechnologieForschungs‐und Entwicklungsgesellschaft mbH an der Universit?t;

    Institut für Diabetes‐TechnologieForschungs‐und Entwicklungsgesellschaft mbH an der Universit?t;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 儿科学;
  • 关键词

    infusion pumps; insulin; insulin infusion systems;

    机译:输液泵;胰岛素;胰岛素输液系统;