首页> 外文期刊>Neurocritical care >Comparison of a Continuous Noninvasive Temperature to Monitor Core Temperature Measures During Targeted Temperature Management
【24h】

Comparison of a Continuous Noninvasive Temperature to Monitor Core Temperature Measures During Targeted Temperature Management

机译:持续无侵入温度的比较监测有针对性温度管理期间核心温度测量的

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Background Temperature modulating devices (TMD) currently utilize core temperature measurements during targeted temperature management (TTM) that are currently limited to esophageal (Et), bladder (Bt), or rectal (Rt) temperatures. We assessed the ability of a continuous noninvasive temperature monitor to accurately approximate core temperature during TTM. Methods All patients undergoing TTM using a gel pad surface TMD and an existing core temperature monitoring device were eligible for this study. Core and continuous noninvasive temperature monitoring values were simultaneously recorded for up to 72 h of TTM. The two sets of temperature data were downloaded from a clinical data acquisition storage system at 1-min intervals. The Bland-Altman method assessed agreement between the core and continuous noninvasive temperature monitor values, by measuring the mean difference (+/- 2 SD) between these values. Results There were 20 subjects that underwent study between January 2018 and March 2018 (55% women, age: 57 +/- 14 years old, BMI: 28.9 + 9.8 kg/m(2), 100% mechanically ventilated). The comparison patient temperature source was predominantly esophageal (n = 10) followed by bladder (n = 5) or rectal (n = 5). There were a total of 999 h of paired patient temperature data from esophageal (50%), bladder (25%), and rectal (25%) temperatures. Bland-Altman analysis demonstrated good agreement with the superficial temperature monitor and core temperature measures in all patients overall, with a difference mean of 0.06 +/- 0.39 C (P = 0.99) and no proportional bias noted (beta =0.002,P = 0.917). Conclusions Continuous noninvasive temperature monitoring is a suitable alternative method for assessing core temperature during TTM. Future studies should focus on developing connectivity with a continuous noninvasive temperature monitor to approximate core temperature during TTM.
机译:None

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号