首页> 外文期刊>European journal of clinical microbiology and infectious diseases: Official publication of the European Society of Clinical Microbiology >Reproducible measurement of vancomycin MICs within the susceptible range in Staphylococcus aureus by a broth microdilution method with a “quasi-continuum” gradient of antibiotic concentrations
【24h】

Reproducible measurement of vancomycin MICs within the susceptible range in Staphylococcus aureus by a broth microdilution method with a “quasi-continuum” gradient of antibiotic concentrations

机译:用肉汤微量稀释法用“准连续素”抗生素浓度的肉毒杆菌易感范围内易感范围内的易感测量

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Abstract The availability of reproducible broth microdilution (BMD) methods including inter log 2 antibiotic dilutions for measuring Staphylococcus aureus (SA) vancomycin minimum inhibitory concentrations (MICs) within the susceptible range is needed to elucidate the impact of vancomycin MICs on clinical outcomes of invasive SA infections. Here, we report on the development of a very precise BMD method that incorporates the following incremental antibiotic concentrations: 0.50, 0.62, 0.75, 0.87, 1.0, 1.25, 1.40, 1.50, 1.60, 1.75, and 2.0?μg/mL. The intra- and inter-assay coefficients of variation of this method were around 20%. The mean of the differences in MIC values for all isolates obtained across two independent runs performed at one center was 0.04?μg/mL [95% confidence interval (CI), 0.011–0.07?μg/mL] and that for ten isolates measured at two different centers was 0.04?μg/mL (95% CI, 0–13?μg/mL). Vancomycin MIC values differed by less than 0.1?μg/mL between runs for most isolates. Storage of isolates at ?20?°C for up to 3?months had no impact on the vancomycin MIC values. The mean vancomycin MIC values obtained by the Etest using a standard inoculum (0.5 McFarland) were significantly higher ( p ?≤ 0.001) than those measured by BMD and the MIC values measured by the two methods correlated poorly (Rho, 0.319; p ?=?0.148). Nevertheless, the mean MIC values measured by the Etest using lower inocula (10 7 or 10 6 ?CFU/mL) and those measured by BMD were comparable and correlated significantly ( p ?=?0.004 for 10 7 ?CFU/mL and p ?=?0.029 for 10 6 ?CFU/mL).
机译:None

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号