首页> 外文期刊>Biomedical Chromatography: An International Journal Devoted to Research in Chromatographic Methodologies and Their Applications in the Biosciences >Bias reduction in the quantitative analysis of a target analyte present in a limited quantity in human plasma using dual-mode heart-cutting two-dimensional liquid chromatography coupled with isotope dilution mass spectrometry
【24h】

Bias reduction in the quantitative analysis of a target analyte present in a limited quantity in human plasma using dual-mode heart-cutting two-dimensional liquid chromatography coupled with isotope dilution mass spectrometry

机译:使用双模心切割二维液相色谱法与同位素稀释质谱法偶联的人血浆中存在有限量的靶分析物的定量分析的定量分析

获取原文
获取原文并翻译 | 示例

摘要

Dual-mode heart-cutting two-dimensional liquid chromatography (DMHC 2D-LC) was applied to isotope dilution mass spectrometry (IDMS) to reduce the bias in the quantitative analysis of a target analyte present in a limited quantity in human plasma. Based on a Waters I-Class LC system, the DMHC 2D-LC system was operated in one- and two-dimensional modes to facilitate the determination of heart-cutting time and the efficient trapping of the target LC eluate. Experiments to determine the feasibility of coupling with IDMS were performed with triple quadrupole mass spectrometry using folic acid standards and/or ~13C_5-folic acid. To validate the performance of the DMHC 2D-LC/IDMS system on a complex sample, human plasma was analyzed for folic acid and the result was compared with that obtained using conventional single-column LC. The total run time of the DMHC 2D-LC system was 20 min, the same as that of the single-column LC system. The peak profile of the spiked ~13C_5-folic acid obtained with single-column LC/MS was affected by matrix effects, but resolved with DMHC 2D-LC/MS, thus improving the accuracy of the analysis. The DMHC 2D-LC/IDMS system showed reliable performance in analyzing the target analyte in human plasma, eliminating matrix effects and saving analysis time.
机译:将双模心切割二维液相色谱(DMHC 2D-LC)施加到同位素稀释质谱(IDMS)上,以减少在人血浆中有限的靶分析物的定量分析中的定量分析中的偏差。基于Waters I-Class LC系统,DMHC 2D-LC系统以一维模式操作,以便于确定心脏切割时间和目标LC洗脱液的有效捕获。用叶酸标准物和/或〜13C_5-叶酸进行三重四极杆质谱法进行偶联与IDMS偶联的可行性的实验。为了验证在复杂样品上的DMHC 2D-LC / IDMS系统的性能,对叶酸分析人血浆,并将结果与​​使用常规单柱LC获得的结果进行比较。 DMHC 2D-LC系统的总运行时间为20分钟,与单列LC系统相同。用单柱LC / MS获得的尖峰〜13C_5-叶酸的峰谱效应受基质效应的影响,但用DMHC 2D-LC / MS分辨,从而提高了分析的准确性。 DMHC 2D-LC / IDMS系统在分析人血浆中的目标分析物中显示可靠性,消除矩阵效应和节省分析时间。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号