Multi-center study for the evaluation of clinical usefulness of attenuation and scatter correction on 201Tl myocardial SPECT

机译:201T1心肌SPECT衰减临床用途评估的多中心研究

获取原文
获取原文并翻译 | 示例
       

摘要

The aim of this study was to evaluate the clinical usefulness of attenuation and scatter correction (AC, SC) on a 201Tl myocardial single-photon emission computed tomography (201Tl SPECT) as a multi-center trial. With a dual-detecter and a triple-detector SPECT systems with a 99mTc transmission source, simultaneous transmission/emission tomography (TCT/ECT) was performed on 38 patients with angiographically coronary heart disease (CHD) and 26 patients without evidence of CHD. Stress and delayed attenuation and scatter corrected images (SAC) and uncorrected images (NC) were reconstructed. On NC images of normal cases, influence of attenuation was greater in male than female. In comparison of 201Tl distribution between male and female, significant decrease in 201Tl activity was observed in the inferoposterior wall in male and that was observed in the anterobasal wall of the left myocardium in female. Such a difference in 201Tl distribution between male and female disappeared on SAC images. On the diagnostic performance for the identification of CHD, SAC images demonstrated improved specificity and accuracy values in the right coronary arterial territory (RCA) with visual analysis statistically. Sensitivity value in the RCA was also improved, but it was not statistically significant. Sensitivity value in the left circumflex arterial territory (LCX) increased without decrease in specificity value on SAC images. In the left anterior descending arterial territory (LAD), sensitivity value increased on SAC images. Although specificity value decreased on SAC images in LAD territory, it was not statistically significant. The difference in 201Tl distribution between male and female is improved in normal cases by attenuation and scatter correction on 201Tl myocardial SPECT. Diagnostic performance of CHD is also improved by attenuation and scatter correction, especially in territories of which specificity in assessing the absence of disease have been suboptimal. In conclusion, attenuation and scatter correction on 201Tl myocardial SPECT is considered to be clinically useful.
机译:本研究的目的是评估衰减和散射校正(AC,SC)对201TL心肌单光子发射计算机断层摄影(201TL SPECT)作为多中心试验的临床有用性。通过双探测器和具有99MTC传输源的三维探测器SPECT系统,在38例患有血管冠心病(CHD)和26名患者的患者上进行同时发送/发射断层扫描(TCT / ECT),没有CHD的证据。重建应力和延迟衰减和散射校正图像(SAC)和未校正图像(NC)。在正常情况的NC图像上,衰减的影响比女性更大。相比,男性和雌性之间的201T1分布,在雄性左侧壁中观察到201T1活性的显着降低,并且在雌性左侧心肌的翼片壁中观察到。在囊图像上消失的男性和女性之间的201Tl分布的这种差异。关于CHD鉴定的诊断性能,SAC图像在右侧分析统计分析中表现出右冠状动脉区域(RCA)中的改善的特异性和准确性值。 RCA中的敏感性值也得到了改善,但它没有统计学意义。左环形动脉区域(LCX)中的灵敏度值增加而没有降低SAC图像的特异性值。在左前期下降动脉域(LAD)中,SAC图像上的灵敏度值增加。虽然在LAD领土上的SAC图像上减少了特异性值,但它没有统计学意义。通过衰减和散射校正在201T1心肌Spect上的正常情况下,在正常情况下改善了201T1分布的差异。 CHD的诊断性能也通过衰减和分散校正来提高,特别是在评估疾病缺失的特异性的领土上是次优的。总之,201T1心肌SPECT上的衰减和散射校正被认为是临床上有用的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号