首页> 外文期刊>Journal of Endodontics: Official Journal of American Association of Endodontists >Detection of Simulated Vertical Root Fractures: Which Cone-beam Computed Tomographic System Is the Most Accurate?
【24h】

Detection of Simulated Vertical Root Fractures: Which Cone-beam Computed Tomographic System Is the Most Accurate?

机译:检测模拟垂直根骨折:哪种锥形光束计算机断层系统最准确?

获取原文
获取原文并翻译 | 示例
       

摘要

Introduction: We aimed to compare the diagnostic accuracy of 5 cone-beam computed tomographic(CBCT) systems in detecting vertical root fractures (VRFs) and to assess whether fracture identification is affected by the presence of root canal filling. Methods: Eighty extracted posterior teeth were included in this study. They were grouped according to the presence/absence of VRFs and the presence/absence of endodontic treatment. The teeth were then inserted in 5 dry skull/mandible assemblies. CBCT scans were performed using 5 different commercially available systems. Two observers evaluated the resultant multiplanar images twice for VRFs using a 3-point scale. Results: i-CAT (Imaging Sciences International, Hatfield, PA) showed the highest diagnostic accuracy in the detection of VRFs among the 5 investigated CBCT systems. The presence of root canal filling did not significantly decrease the ability to detect VRFs by all the studied systems. In the detection of VRFs in endodontically treated teeth, i-CAT was the most accurate, whereas 3D Accuitomo (J. Morita, Kyoto, Japan) was the least. Interobserver agreement was moderate for the i-CAT and fair for the rest of the studied modalities, whereas intraobserver agreement was good for the Scanora 3D (Soredex, Tuusula, Finland) and moderate for images from the other CBCT machines. Conclusions: At the specified exposure parameters in the detection of VRFs in non root canal filled teeth, i-CAT showed the highest diagnostic accuracy followed by Planmeca Promax 3D (Planmeca, Helsinki, Finland), Scanora 3D, Accuitomo 3D, and Galileos 3D (Sirona Dental Systems, Bensheim, Germany), respectively. In the detection of VRFs in root canal filled teeth, i-CAT showed the highest diagnostic accuracy followed by Planmeca Promax 3D, Scanora 3D, Galileos Comfort (Sirona Dental Systems), and Accuitomo 3D.
机译:简介:我们旨在比较5锥形束计算机断层摄影(CBCT)系统检测垂直根骨折(VRF)的诊断精度,并评估裂缝鉴定是否受根管填充的存在影响。方法:本研究包括八十次提取的后牙。它们根据VRF的存在/不存在进行分组和牙髓治疗的存在/不存在。然后将牙齿插入5个干颅骨/下颌骨组件中。 CBCT扫描使用5种不同的市售系统进行。两个观察者使用3点刻度评估由3点刻度进行两次VRF的Multiplanar图像。结果:I-Cat(映像科学国际,Hatfield,PA)在5调查的CBCT系统中检测VRF的检测中显示出最高的诊断准确性。根管填充的存在没有显着降低所有研究的系统检测VRF的能力。在脊髓牙齿上的VRF检测中,I-CAT是最准确的,而3D accuitomo(J. Morita,京都,日本)是最重要的。 Interobserver协议为I-Cat的适度和公平的学习模式,而奥里诺·美国机构协议对于Scanora 3D(SoreDex,Tuusula,Finland)对来自其他CBCT机器的图像有益。结论:在检测中的指定曝光参数,在填充非根管填充牙齿的VRF中,I-CAT显示出最高的诊断准确性,然后是Planmeca Promax 3D(Planmeca,赫尔辛基,芬兰),Scanora 3D,返回术语3D和Galileos 3D( Sirona牙科系统,Bensheim,德国)分别。在检测到根管填充牙齿中的VRF中,I-Cat展示了最高的诊断精度,然后是Planmeca Promax 3D,Scanora 3D,伽利冬舒适(Sirona牙科系统)和返回术。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号