首页> 外文期刊>The Journal of hand surgery, European volume >High revision rates with the metal-on-metal Motec carpometacarpal joint prosthesis
【24h】

High revision rates with the metal-on-metal Motec carpometacarpal joint prosthesis

机译:高修订率与金属型Motec CarpometCarpal关节假肢

获取原文
获取原文并翻译 | 示例
           

摘要

We retrospectively evaluated a consecutive series of 42 Motec thumb carpometacarpal total joint arthroplasties. The primary endpoint was revision with implant removal and trapeziectomy. At follow-up the disability of the arm shoulder and hand (DASH) score, pain on numerical rating scale at rest and with activity and serum chrome and cobalt concentrations were assessed for both unrevised and revised patients. At a mean follow-up of 26 months, 17 patients had been revised. The 2 year cumulative revision rate was 42% (95% CI, 28-60%). The DASH score and pain scores at rest and with activity were comparable between the patients whose thumbs remained unrevised and those revised. Patients with elevated serum chrome and cobalt levels had significantly higher DASH and pain scores, but elevated levels were not associated with revision. The revision rate in this study is unacceptably high. However, pain and DASH scores after revision are acceptable and comparable with patients with non-revised implants.
机译:我们回顾性地评估了连续的42个Motec Thumb Carpometacal arpal总关节关节塑料。主要终点是具有植入物去除和梯形切除术的修正。在跟踪臂肩部和手(跳跃)的残疾时,评估休息和血清铬和血清铬和钴浓度的数值评级规模的疼痛。在26个月的平均随访时,17名患者已修改。 2年累积修正率为42%(95%CI,28-60%)。休息时间和活动的损伤分数和疼痛评分与患者之间的患者相比,拇指仍未观察并修订的患者。血清铬和钴水平升高的患者具有显着提高的瘫痪和疼痛评分,但水平升高无关。该研究的修订率是不可接受的。然而,修改后疼痛和划伤分数是可接受的,并且与未修订植入物的患者相当。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号