Performance of a linear accelerator couch positioning quality control task using an electronic portal imaging device

机译:使用电子门户成像装置的线性加速器沙发定位质量控制任务的性能

获取原文
获取原文并翻译 | 示例
           

摘要

Short and semi-automated quality assurance (QA) programs are becoming one of the most popular and highly demanding tasks in radiotherapy. The current research investigates the accuracy of a four degrees of freedom (4DoF) medical linear accelerator couch positioning with a fast and accurate method based on images acquired using an electronic portal imaging device (EPID). An accurate EPID QA phantom and a proper in-house code were used. A Siemens medical linear accelerator equipped with an a-Si EPID was used to acquire portal images. For verifying the mechanical performance of the EPID positioning, EPID sensitivity, and accuracy of the code response from the image processing point of view were investigated. To characterize the results, three deviations in the phantom positioning were deliberately created. The translational and rotational displacements of the treatment couch were then evaluated. The loading effect on the treatment couch was then investigated. The results of prerequisite tests, including the mechanical performance of the EPID, and the sensitivity and accuracy of the recognition codes, were assessed. The results were found to be within the tolerance range reported at AAPM TG-142. The mean deviations of the tests between expected and measured displacements by 4DoF treatment couch were found to be 0.13 degrees +/- 0.11 degrees, 0.12 +/- 0.17 mm, 0.17 +/- 0.13 mm, and 0.04 +/- 0.09 mm for rotational, longitudinal, lateral, and vertical shifts, respectively. The results showed that the proposed method is a reliable and fast approach to find the uncertainties occurring intreatment couch positioning.
机译:短期和半自动质量保证(QA)计划正在成为放射疗法中最受欢迎和最苛刻的任务之一。目前的研究调查了通过使用电子门户成像装置(EPID)获取的图像的快速和准确的方法,研究了四个自由度(4dof)医用线性加速器沙发定位的准确性。使用精确的EPID QA幻像和适当的内部代码。使用配备A-SI EPID的西门子医用线性加速器来获取门户网站图像。为了验证EPID定位的机械性能,EPID敏感性和从图像处理的图像处理的代码响应的准确性进行了研究。为了表征结果,故意创建幻像定位中的三个偏差。然后评估治疗沙发的平移和旋转位移。然后研究了处理沙发上的负载效果。评估先决条件测试的结果,包括EPID的机械性能以及识别码的灵敏度和准确性。发现结果在AAPM TG-142报道的耐受范围内。 4DOF处理沙发上预期和测量位移之间的测试的平均偏差为0.13度+/- 0.11度,0.12 +/- 0.17 mm,0.17 +/- 0.13mm,0.04 +/- 0.09 mm,用于旋转分别,纵向,横向和垂直移位。结果表明,该方法是一种可靠而快速的方法,可以找到发生的髂内沙发定位的不确定性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号