Deep brain stimulation (DBS) efficacy is related to optimal electrode placement. Several authors have quantified brain shift related to surgical targeting; yet, few reports document and discuss the effects of brain shift after insertion. Objective: To quantify brain shift and electrode displacement after device insertion. Twelve patients were retrospectively reviewed, and one post-operative MRI and one time-delayed CT were obtained for each patient and their implanted electrodes modeled in 3D. Two competing methods were employed to measure the electrode tip location and deviation from the prototypical linear implant after the resolution of acute surgical changes, such as brain shift and pneumoceph-alus. In the interim between surgery and a pneumocephalus free postoperative scan, electrode deviation was documented in all patients and all electrodes. Significant shift of the electrode tip was identified in rostral, anterior, and medial directions (p < 0.05). Shift was greatest in the rostral direction, measuring an average of 1.41 mm. Brain shift and subsequent electrode displacement occurs in patients after DBS surgery with the reversal of intraoperative brain shift. Rostral displacement is on the order of the height of one DBS contact. Further investigation into the time course of intraoperative brain shift and its potential effects on procedures performed with rigid and non-rigid devices in supine and semi-sitting surgical positions is needed.
机译:深脑刺激(DBS)功效与最佳电极放置有关。若干作者已经量化了与手术靶向有关的脑移位;然而,很少有报道文件,并讨论插入后脑移位的影响。目的:通过装置插入后量化脑换档和电极位移。回顾性地回顾了12名患者,并为每位患者及其在3D中建模的植入电极获得了一个术后MRI和一次延迟CT。采用两种竞争方法测量电极尖端位置并在分辨急性手术变化之后从原型线性植入物偏离,例如脑移和肺肺肺血糖。在手术和肺术术后扫描之间的临时,在所有患者和所有电极中记录了电极偏差。电极尖端的显着偏移在讲轴承,前部和内侧方向上识别(P <0.05)。换档方向最大,平均尺寸为1.41毫米。 DBS手术后患者发生脑移和后续电极位移,术后脑移位逆转。振动式位移是在一个DBS接触的高度的顺序上。需要进一步调查术中脑移位的时间过程及其对仰卧和半坐着外科手术位置的刚性和非刚性装置进行的潜在影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号