Direct spectral measurements with a Brewer spectroradiometer: absolute calibration and aerosol optical depth retrieval

机译:使用Brewer分光光度计进行直接光谱测量:绝对校准和气溶胶光学深度检索

获取原文
获取原文并翻译 | 示例
           

摘要

We present three different methods for the absolute calibration of direct spectral irradiances measured with a Brewer spectroradiometer, which are shown to agree to within ±2%. Direct irradiance spectra derived by Brewer and Bentham spectroradiometers agree to within 4±3%. Good agreement was also found by a comparison of the aerosol optical depth and Angstrom exponent retrieved by the two instruments and a multifilter rotational shadowband radiometer. The spectral aerosol optical depth (300-365 nm) derived from six years of direct irradiance measurements at Thessaloniki shows a distinct seasonal variation, averaging to ~0.3 at 340 nm in winter and ~0.7 in summer.
机译:我们介绍了三种用于用Brewer分光光度计测量的直接光谱辐照度的绝对校准的不同方法,这些方法显示在±2%的范围内。 Brewer和Bentham分光光度计得出的直接辐射光谱在4±3%之内。通过比较两种仪器和多滤镜旋转阴影带辐射计获得的气溶胶光学深度和埃指数,也发现了很好的一致性。塞萨洛尼基通过六年的直接辐照度测量得出的光谱气溶胶光学深度(300-365 nm)表现出明显的季节变化,冬季平均在340 nm时约为〜0.3,夏季平均为〜0.7。