Multichannel radiometer calibration: a new approach

机译:多通道辐射计校准:一种新方法

获取原文
获取原文并翻译 | 示例
           

摘要

The error in irradiance measured with Sun-calibrated multichannel radiometers may be large when the solar zenith angle (SZA) increases. This could be particularly detrimental in radiometers installed at mid and high latitudes, where SZAs at noon are larger than 50 deg during part of the year. When a multi-regressive methodology, including the total ozone column and SZA, was applied in the calculation of the calibration constant, an important improvement was observed. By combining two different equations, an improvement was obtained at almost all the SZAs in the calibration. An independent test that compared the irradiance of a multichannel instrument and a spectroradiometer installed in Ushuaia, Argentina, was used to confirm the results.
机译:当太阳天顶角(SZA)增大时,使用经过Sun校准的多通道辐射计测量的辐照度误差可能会很大。这对于中高纬度地区安装的辐射计尤其有害,在该年中,中午SZA大于50度。当使用包括总臭氧柱和SZA在内的多元回归方法来计算校准常数时,观察到了重要的改进。通过组合两个不同的方程,在校准中几乎所有SZA上都得到了改善。一项独立测试将多通道仪器和安装在阿根廷乌斯怀亚的光谱辐射仪的辐照度进行了比较,以确认结果。