A modified technique to minimize filter tilting during deployment of the Gunther Tulip filter: in vitro study.

机译:一种改进的技术,可在使用Gunther Tulip过滤器时最大程度地减少过滤器倾斜:体外研究。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

With an in vitro model of the inferior vena cava, it was demonstrated that slight tension on the introducer during quick release of the jugular Gunther Tulip filter led to less tilting of the filter compared with the technique recommended by the manufacturer. This technique may be useful to minimize significant filter tilting that may be associated with decreased filtration efficiency, and difficulty or impossibility of future filter retrieval.
机译:在下腔静脉的体外模型中,已证明与制造商推荐的技术相比,在快速释放颈状Gunther Tulip过滤器的过程中,导引器上的轻微张紧导致过滤器的倾斜较小。该技术可用于使可能与降低的过滤效率以及将来的过滤器回收的困难或不可能相关的明显的过滤器倾斜最小化。