This article describes a fully automatic, real-time, freehand ultrasound calibration system. The system was designed to be simple and sterilizable, intended for operating-room usage. The calibration system employed an automatic-error-retrieval and accuracy-control mechanism based on a set of ground-truth data. Extensive validations were conducted on a data set of 10,000 images in 50 independent calibration trials to thoroughly investigate the accuracy, robustness, and performance of the calibration system. On average, the calibration accuracy (measured in three-dimensional reconstruction error against a known ground truth) of all 50 trials was 0.66 mm. In addition, the calibration errors converged to submillimeter in 98% of all trials within 12.5 s on average. Overall, the calibration system was able to consistently, efficiently and robustly achieve high calibration accuracy with real-time performance.
机译:本文介绍了一种全自动,实时,徒手的超声校准系统。该系统设计简单,可消毒,适用于手术室使用。该校准系统采用了基于一组真实数据的自动错误检索和精度控制机制。在50个独立的校准试验中,对10,000张图像的数据集进行了广泛的验证,以彻底研究校准系统的准确性,鲁棒性和性能。平均而言,所有50个试验的校准精度(针对已知的地面真实情况,通过三维重建误差测量)为0.66 mm。此外,平均所有校准中的98%的校准误差平均在12.5 s内收敛到亚毫米级。总体而言,校准系统能够始终如一地,高效且稳健地实现具有实时性能的高校准精度。