Abstract. To manage fractures in long bones, intramedullar implants, plates or external fixators are often used. In many cases,nthe implants are removed after bone consolidation. X-ray images are normally used to monitor bone formation and to determinenthe point of return to full load bearing and removal of the implants. However, plain radiographs give only inaccurate informationnabout the degree of healing progress. Known quantitative methods as QCT, DEXA, etc. provide information about bone densitynwhich certainly contributes to the mechanical properties of healing bone, but they do not provide a direct measurement of thenstiffness of the healing callus. In this study we present an in vivo 4-point-bending stiffness device for small animals which isndesigned to directly monitor the progression of the healing process. The device was tested in a bone-defect model with differentntest-specimens chosen to simulate the stiffness of bone at different stages of healing. To verify the results, it was tested in annanimal fracture study in rabbits during the healing period with and without an intramedulary implant. Both the test-specimennand bones of the in vivo study were compared with data in a materials testing system (MTS) in four-point bending. The devicenwas found to have a high precision and significant in vitro and in vivo correlation with the MTS. The results suggest that thisnmeasurement device has the ability to monitor the healing process of bone and to analyse the influence of degradable implantsnon the mechanical behaviour of bone or bone metabolism effecting pharmaceutics.
机译:抽象。为了处理长骨骨折,通常使用髓内植入物,钢板或外固定器。在许多情况下,在骨固结后不去除植入物。 X射线图像通常用于监视骨形成并确定返回到满负荷和植入物取出的位置。但是,普通的X光片只能提供有关治疗进展程度的不准确信息。已知的定量方法(如QCT,DEXA等)提供了有关骨密度的信息,这些信息无疑有助于愈合骨的机械性能,但不能直接测量愈合骨call的刚度。在这项研究中,我们提出了一种用于小动物的体内四点弯曲刚度装置,该装置旨在直接监测愈合过程的进展。该设备在骨骼缺陷模型中进行了测试,并选择了不同的试样来模拟不同愈合阶段的骨骼刚度。为了验证结果,在有或没有髓内植入物的愈合期的兔子的骨折试验中对它进行了测试。将体内研究的测试样本和骨骼与材料测试系统(MTS)中四点弯曲的数据进行了比较。已经发现该装置具有高精度,并且与MTS在体外和体内具有显着相关性。结果表明,该测量装置具有监测骨骼愈合过程和分析可降解植入物的影响的能力,而骨骼的力学行为或影响药物代谢的骨骼代谢均不受影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号