首页> 外文期刊>IEEE sensors journal >High-Precision Humidity Sensor Response Time Measurement Based on the Switchable Chamber With the Hydrophobic Interface
【24h】

High-Precision Humidity Sensor Response Time Measurement Based on the Switchable Chamber With the Hydrophobic Interface

机译:高精度湿度传感器响应时间测量基于可切换腔室的疏水界面

获取原文
获取原文并翻译 | 示例
           

摘要

A high-precision humidity response time measurement has been urgently needed for the fast response humidity sensor in some special situations, for instance, when the quadcopter performs the meteorological observations in the atmospheric boundary layer, the relative humidity changes drastically as it flies into the cloud. So it is essential to gain the accurate response time for these sensors. However, owing to the uncontrollable errors, the conventional methods are difficult to carry out an accurate measurement. Herein, for meeting the requirements above, we designed an accurate measurement method based on a type of switchable chamber and a humidity generator. Among, the chamber was designed through surface molecular dynamics analysis and verified by experiments. It is proved that the switchable chamber with the hydrophobic surface and low water molecule content, combined with the surface to volume ratio S/V = 0.2 can achieve the realistic humidity response time of the sensor. For a commercial sensor (SHT20) at 25 degrees C, the time corresponding to the 63% and 90% complete responses are 7 s and 49 s measured in this way, which is extraordinarily improved than those traditional ways (51 s and 140 s). As a result, this measurement method could gain the accurate response time for the special situation applications.
机译:在一些特殊情况下,迫切需要高精度湿度响应时间测量,例如,当Quadcopter在大气边界层中的气象观测中进行气象观测时,相对湿度随着它的飞向云而变化。因此,必须为这些传感器获得准确的响应时间。然而,由于不可控制的误差,传统方法难以进行准确测量。这里,为了满足上述要求,我们设计了一种基于一种可切换腔室和湿度发生器的精确测量方法。其中,腔室通过表面分子动力学分析设计并通过实验验证。证明,可切换室具有疏水性表面和低水分子含量,与表面与体积比S / V = 0.2组合可以实现传感器的逼真湿度响应时间。对于25摄氏度的商业传感器(SHT20),对应于63%和90%的完全响应的时间是以这种方式测量的7 s和49s,这比传统方式(51 s和140s)非常改善。结果,该测量方法可以获得特殊情况应用的准确响应时间。

著录项

  • 来源
    《IEEE sensors journal》 |2020年第10期|5566-5574|共9页
  • 作者单位

    Chinese Acad Sci State Key Lab Transducer Technol Aerosp Informat Res Inst Beijing 100190 Peoples R China|Univ Chinese Acad Sci Sch Elect Elect & Commun Engn Beijing 100049 Peoples R China;

    Chinese Acad Sci State Key Lab Transducer Technol Aerosp Informat Res Inst Beijing 100190 Peoples R China|Univ Chinese Acad Sci Sch Elect Elect & Commun Engn Beijing 100049 Peoples R China;

    Univ Chinese Acad Sci Sch Elect Elect & Commun Engn Beijing 100049 Peoples R China;

    Chinese Acad Sci State Key Lab Transducer Technol Aerosp Informat Res Inst Beijing 100190 Peoples R China|Univ Chinese Acad Sci Sch Elect Elect & Commun Engn Beijing 100049 Peoples R China;

    Chinese Acad Sci State Key Lab Transducer Technol Aerosp Informat Res Inst Beijing 100190 Peoples R China;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Humidity; Switches; Temperature measurement; Temperature sensors; Adsorption; Time measurement; Humidity response time measurement; switchable chamber; high-precision; hydrophobic interface; fast response;

    机译:湿度;开关;温度测量;温度传感器;吸附;时间测量;湿度响应时间测量;可切换室;高精度;疏水界面;快速响应;