首页> 外文期刊>Sensors and Actuators >Electrochemical paper-based analytical device for multiplexed, point-of-care detection of cardiovascular disease biomarkers
【24h】

Electrochemical paper-based analytical device for multiplexed, point-of-care detection of cardiovascular disease biomarkers

机译:基于电化学纸的分析装置,用于复用,心血管疾病生物标志物的护理点检测

获取原文
获取原文并翻译 | 示例
           

摘要

Cardiovascular diseases (CVDs) are the number one cause of death globally. Therefore, a sensitive and affordable point-of-care sensing device is needed for the diagnosis and monitoring of diseases. Herein, we demonstrated an electrochemical paper-based analytical device (ePAD) coupled with a label-free immunoassay for simultaneous determination of three important CVDs biomarkers including C-reactive protein (CRP), troponin I (cTnl), and procalcitonin (PCT). The ePAD was created by stacking a wax-patterned paper and a transparency film using laser-cut double-sided adhesive to define the sample inlet, three separate detection zones for each analyte, and fluidic channels that connect the inlet and detection zones. Antibodies to the target biomarkers were immobilized on graphene oxide (GO)-modified carbon electrodes stencil-printed on the ePAD. Concentrations of the CVDs biomarkers were determined by square wave voltammetry (SWV). A significant decrease in current response in a concentration-dependent manner was observed in the presence of the cardiac biomarkers while there was no observable change in current response in the absence of the biomarkers. Good linearity (R~2 > 0.99), low detection limits (0.38 ng mL~(-1), 0.16 pg mlL~(-1), and 0.27 pg mL~(-1) for CRP, cTnI, and PCT, respectively), and good reproducibility (RSD < 5 %) were achieved for the quantifications. Furthermore, the immunosensor has been applied to simultaneously detect the three biomarkers in serum samples with satisfactory results. The proposed immunosensor can be a great alternative for the early detection of cardiovascular diseases at the point-of-care.
机译:心血管疾病(CVDS)是全球死亡的第一原因。因此,诊断和监测疾病需要敏感和实惠的护理点感测装置。在此,我们证明了一种基于电化学纸的分析装置(EPAD),其与无标记的免疫测定结合,用于同时测定包括C反应蛋白(CRP),肌钙蛋白I(CTN1)和ProCalcitonin(PCT)的三个重要的CVDS生物标志物。通过使用激光剪切双面粘合剂堆叠蜡状纸和透明胶片来创建EPAD,以限定样品入口,每个分析物的三个单独的检测区,以及连接入口和检测区域的流体通道。将目标生物标志物的抗体固定在石墨烯氧化物(GO)制片的碳电极上固定在EPAD上印刷。通过方波伏安法(SWV)测定CVDS生物标志物的浓度。在心脏生物标志物存在下观察到以浓度依赖性方式的电流响应的显着降低,而在没有生物标志物的情况下,目前响应没有可观察到的变化。良好的线性度(R〜2> 0.99),低检测限(0.38ng ml〜(-1),0.16pg mll〜(-1),CRP,CTNI和PCT的0.27pg ml〜(-1), ),对量化实现良好的再现性(RSD <5%)。此外,已施用免疫传感器以在血清样品中同时检测三种生物标志物,结果令人满意。所提出的免疫传感器可以是在护理点的早期检测心血管疾病的替代方案。

著录项

  • 来源
    《Sensors and Actuators》 |2021年第3期|129336.1-129336.8|共8页
  • 作者单位

    Electrochemistry and Optical Spectroscopy Center of Excellence (EOSCE) Department of Chemistry Chulalongkorn University Pathumwan Bangkok 10330 Thailand;

    Department of Chemistry Colorado State University Fort Collins CO 80523 USA Institute of Nano Science and Technology Hanyang University Seoul 04763 South Korea;

    Department of Chemistry Colorado State University Fort Collins CO 80523 USA Department of Pharmaceutical Chemistry Faculty of Pharmacy Universitas Gadjah Mada Yogyakarta 55281 Indonesia;

    Department of Chemistry Faculty of Science Srinakharinwirot University Wattana Bangkok 10110 Thailand;

    Electrochemistry and Optical Spectroscopy Center of Excellence (EOSCE) Department of Chemistry Chulalongkorn University Pathumwan Bangkok 10330 Thailand Center of Excellence on Petrochemical and Materials Technology Chulalongkorn University Pathumwan Bangkok 10330 Thailand;

    Department of Chemistry Colorado State University Fort Collins CO 80523 USA;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Paper-based analytical devices; Point-of-care detection; Multiplexed electrochemical detection; Cardiovascular disease biomarkers;

    机译:基于纸张的分析装置;护理点检测;多路复用电化学检测;心血管疾病生物标志物;