Effects of neutron energy spectrum on the efficiency calibration of epithermal neutron detectors

机译:中子能谱对超热中子探测器效率标定的影响

获取原文
           

摘要

The 235U epithermal neutron fission detector system at TFTR has been calibrated in situ, using three different neutron sources: 252Cf, a D‐D, and a D‐T neutron generator. These sources produced a continuous spectrum, a line spectrum at ∼2.5 MeV, and a line spectrum at ∼14 MeV, respectively. The toroidally averaged detection efficiencies are 4.61×10-9 (252Cf), 4.80×10-9 (D‐D), and 5.42×10-9 (D‐T) countseutron for a moderated 1.3‐g 235U detector placed ∼4 m from the torus centerline. The effects of (n,2n) and (n,n’γ) reactions are discussed.
机译:TFTR的235U超热中子裂变探测器系统已经使用三种不同的中子源进行了现场校准:252Cf,D‐D和D‐T中子发生器。这些源分别产生连续谱,〜2.5 MeV的线谱和〜14 MeV的线谱。放置在1.3g 235U缓和检测器中的中子的环形平均检测效率为4.61×10-9(252Cf),4.80×10-9(D‐D)和5.42×10-9(D‐T)计数/中子〜距圆环中心线4 m。讨论了(n,2n)和(n,n’γ)反应的影响。