A Scanning Michelson Interferometer for the Measurement of the Concentration and Temperature Derivative of the Refractive Index of Liquids

机译:扫描迈克尔逊干涉仪,用于测量液体折射率的浓度和温度导数

获取原文
获取原文并翻译 | 示例
       

摘要

A compact scanning Michelson interferometer for the measurement of both the concentration and temperature derivative of the refractive index of liquids, liquid mixtures, and solutions has been developed. Scanning of one of the mirrors allows the extraction of the phase information and makes the instrument insensitive to intensity drifts of the light source. The refractive index noise level of the empty instrument is of the order of 10~(-8) for a 10 mm sample cell, and for continuously varying refractive indices there is no upper limit of the linear detection range. Accuracies of the order of 0.5% can be achieved both for (((partial deriv)n)/((partial deriv)c))_(p,T) and (((partial deriv)n)/((partial deriv)T))_(p,c).
机译:已经开发出用于测量液体,液体混合物和溶液的折射率的浓度和温度导数的紧凑型扫描迈克尔逊干涉仪。扫描其中一面镜子可以提取相位信息,并使仪器对光源的强度漂移不敏感。空仪器的折射率噪声水平对于10 mm样品池为10〜(-8)数量级,对于连续变化的折射率,线性检测范围没有上限。对于(((偏导数)n)/((偏导数c))_(p,T)和(((偏导数n)/((偏导数)))都可以达到0.5%的精度T))_(p,c)。