首页> 外文期刊>Radiation Protection Dosimetry >THE ESTABILISTMENT OF REFERENCE DOSTES IN PAEDIATRIC RADIOLOGY AS A FUNCTION OF PATIENT SIZE
【24h】

THE ESTABILISTMENT OF REFERENCE DOSTES IN PAEDIATRIC RADIOLOGY AS A FUNCTION OF PATIENT SIZE

机译:小儿放射学中参考剂量的建立作为患者大小的作用

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

There is a wide range in paediatric patient size from a newborn baby tio a 15-year-old adolescent. Reference doses for paediatric radiology can sensibly be established only for specific sizes of children. Five standard sizes of patient have been chosne at ages 0 (newborn), 1,5,10 and 15 years. Standard AP and lateral thicknesses for the head and trunk for the refernece ages wre dervied from publsihed measuremetns on children. Normalisation factors for entrance surfawce dose and dose-area product measurements were calculated which depend on the thickness of the real patient, the thickness of the nearest standard `patient', and an effective linear attenuation coefficient (μ). These normalisation factors were applied to European data to dervie some preliminary reference doses.
机译:从新生婴儿到15岁青少年,儿科患者的大小范围很广。仅针对特定大小的儿童,才可以合理地确定儿科放射学的参考剂量。五种标准尺寸的患者在0(新生儿),1、5、10和15岁时选择了胆囊炎。参考年龄的标准AP和头部和躯干的侧向厚度源自儿童的公开度量。计算了入口冲浪剂量和剂量面积乘积测量值的归一化因子,该归一化因子取决于实际患者的体重,最近的标准“患者”的厚度以及有效的线性衰减系数(μ)。将这些归一化因子应用于欧洲数据,以推算一些初步的参考剂量。