首页> 外文期刊>Radiation Protection Dosimetry >A SURVEY OF 14 COMPUTED TOMOGRAPHY SCANNERS IN GREECE AND 32 SCANNERS IN ITALY. EXAMINATION FREQUENCIES, DOSE REFERENCE VALUES, EFFECTIVE DOSES AND DOSES TO ORGANS
【24h】

A SURVEY OF 14 COMPUTED TOMOGRAPHY SCANNERS IN GREECE AND 32 SCANNERS IN ITALY. EXAMINATION FREQUENCIES, DOSE REFERENCE VALUES, EFFECTIVE DOSES AND DOSES TO ORGANS

机译:对希腊的14台计算机断层扫描仪和意大利的32台扫描仪进行了调查。考试频率,剂量参考值,有效剂量和器官剂量

获取原文
获取原文并翻译 | 示例
       

摘要

A survey of examination frequencies, dose reference values, effective doses and doses to organs involving 14 scanners from Greece and 32 scanners from Italy was carried out for the years 1999 and 2000. Examination frequencies per scanner and per year were found to be 3590 for Greece and 4520 for Italy. For the types of examinations considered, CDTI_w and DLP measurements were taken. Also scan lengths used for the same types of examinations were monitored. For the same types of examinations effective doses were calculated by two methods, and it was found that their mean values ranged from 13.1 mSv for thoracic spine to 1.6 mSv for the brain examinations. From the data of the 14 Greek laboratories, doses to organs were calculated and it was found that the thyroid receives 50.2 +- 19.8 mGy during a cervical spine examination while the gonads receive 17.8 +- 6.9 mGy during a routine pelvis examination.
机译:在1999年和2000年,对希腊的14台扫描仪和意大利的32台扫描仪进行了检查频率,剂量参考值,有效剂量和器官剂量的调查。希腊每台扫描仪和每年的检查频率为3590。意大利为4520。对于考虑的检查类型,进行了CDTI_w和DLP测量。还监视用于相同类型检查的扫描长度。对于相同类型的检查,通过两种方法计算了有效剂量,发现它们的平均值范围从胸椎的13.1 mSv到脑部检查的1.6 mSv。从14个希腊实验室的数据中,计算出器官的剂量,发现在颈椎检查期间甲状腺接受50.2±19.8 mGy,而在常规骨盆检查期间性腺接受17.8±6.9 mGy。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号