ORGAN AND EFFECTIVE DOSE EVALUATION IN DIAGNOSTIC RADIOLOGY BASED ON IN-PHANTOM DOSE MEASUREMENTS WITH NOVEL PHOTODIODE-DOSEMETERS

机译:基于新型光电二极管剂量的剂量内剂量测量的诊断放射学中的器官和有效剂量评估

获取原文
获取原文并翻译 | 示例
           

摘要

Organ and the effective doses of patients undergoing clinical X ray examinations of chest and abdomen were evaluated with an anthropomorphic phantom and a new dosimetry system. The system was comprised of 34 pin photodiode dosemeters placed in/on particular tissues or organs of the anthropomorphic phantom, where the tissues and organs are defined by the International Commission on Radiological Protection (ICRP) to estimate the effective doses. Dosemeter signals were acquired on a personal computer directly, and converted into absorbed doses, from which the organ and the effective doses were evaluated on the computer. Our study showed that organ doses ranged from <0.01 to 0.72 mGy in routine X-ray radiography of chest and of abdomen and from 0.07 to 55.91 mGy in routine computed tomography (CT) examinations with current multi-slice CT scanners. The effective dose observed in the chest CT examination was approximately 300 times higher than that in chest radiography.
机译:用拟人模型和新的剂量测定系统评估了接受胸部和腹部临床X射线检查的患者的器官和有效剂量。该系统由34针光电二极管剂量计组成,该剂量计放置在拟人化体模的特定组织或器官中/上,其中组织和器官由国际放射防护委员会(ICRP)定义以估算有效剂量。剂量计信号直接在个人计算机上获取,并转换为吸收剂量,然后在计算机上评估器官和有效剂量。我们的研究表明,在使用常规多层CT扫描仪进行的常规胸部X线平片X线照相中,器官剂量范围为<0.01至0.72 mGy,在常规计算机断层扫描(CT)检查中,器官剂量范围为0.07至55.91 mGy。在胸部CT检查中观察到的有效剂量比在胸部X光检查中所观察到的有效剂量高大约300倍。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号