首页> 外文期刊>Radiation Protection Dosimetry >RESPONSE OF NEUTRON DETECTORS TO HIGH-ENERGY MIXED RADIATION FIELDS
【24h】

RESPONSE OF NEUTRON DETECTORS TO HIGH-ENERGY MIXED RADIATION FIELDS

机译:中子探测器对高能混合辐射场的响应

获取原文
获取原文并翻译 | 示例

摘要

Radiation protection around CERN's high-energy accelerators represents a major challenge due to the presence of complex, mixed radiation fields. Behind thick shielding neutrons dominate and their energy ranges from fractions of eV to about 1 GeV. In this work the response of various portable detectors sensitive to neutrons was studied at CERN's High-Energy Reference Field Facility (CERF). The measurements were carried out with conventional rem counters, which usually cover neutron energies up to 20 MeV, the Thermo WENDI-2, which is specified to measure neutrons up to several GeV, and a tissue-equivalent proportional counter. The experimentally determined neutron dose equivalent results were compared with Monte Carlo (MC) simulations. Based on these studies field calibration factors can be determined, which result in a more reliable estimate of H~*(10) in an unknown, but presumably similar high-energy field around an accelerator than a calibration factor determined in a radiation field of a reference neutron source.
机译:由于存在复杂的混合辐射场,CERN高能加速器周围的辐射防护成为一项重大挑战。在厚屏蔽中子的背后,能量占主导地位,其能量范围从eV的分数到大约1 GeV。在这项工作中,CERN的高能参考场设施(CERF)研究了各种对中子敏感的便携式探测器的响应。使用常规的雷姆计数器(通常可测量高达20 MeV的中子能量),Thermo WENDI-2(可测量高达几个GeV的中子)和组织等效比例计数器进行测量。将实验确定的中子剂量当量结果与蒙特卡洛(MC)模拟进行比较。基于这些研究,可以确定场校准因子,从而在加速器周围的未知但大概是与加速器周围的高能场相似的情况下,可以更可靠地估算H〜*(10)。参考中子源。