Comparative measurements of the beta-particle depth dose distribution for a planar ~(147)Pm source were determined using an extrapolation chamber (ECH), radiochromic dye films (RDF) and ~7LiF ultra thin thermoluminescence dosemeters (utTLD). The absolute dose estimates provided by the ECH are in excellent agreement with respect to those obtained from RDF dosemeters. In general the agreement between the ECH and the RDF were within 10. The depth doses initially obtained using ~7LiF utTLD dosemeters assuming the dosemeter centre as the effective point of measurement exhibited an overestimation from 18 for the first 10 mg.cm~(-2) up to 60 at 30 mg.cm~(-2) with respect to the ECH measurements. However in this assumption the utTLD estimations neglected the dose gradient inside the dosemeter. In order to take this into account the observed relation between the dose rate measured for each utTLD disc and its corresponding mass thickness was used to calculate the dose rate at the disc surface. A significant improvement in terms of agreement with the absolute measurements was achieved using this approach. The much higher sensitivity of utTLDs compared with ECH and RDF make this technique particularly useful for absorbed dose measurements for low activity beta particle sources.
机译:使用外推室(ECH),放射致变色染料膜(RDF)和〜7LiF超薄热致发光剂量计(utTLD)确定了平面〜(147)Pm源的β粒子深度剂量分布的比较测量结果。 ECH提供的绝对剂量估计值与从RDF剂量计获得的绝对剂量估计值非常一致。通常,ECH和RDF之间的一致性在10以内。假设剂量计中心为有效测量点,使用〜7LiF utTLD剂量计最初获得的深度剂量显示出高估了头18 mg.cm〜(-2 )相对于ECH测量值在30 mg.cm〜(-2)时高达60。但是,在此假设下,utTLD估算忽略了剂量计内部的剂量梯度。为了考虑到这一点,对每个utTLD光盘测量的剂量率与其对应的质量厚度之间的观察关系用于计算光盘表面的剂量率。使用这种方法,在与绝对测量值的一致性方面有了显着改善。与ECH和RDF相比,utTLD的灵敏度高得多,使得该技术特别适用于低活性β颗粒源的吸收剂量测量。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号