RESPONSE OF TLD-300 TO HEAVY CHARGED PARTICLES

机译:TLD-300对带电粒子的响应

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

The TL response of CaF_2:Tm to 0.7 MeV protons, 5.3 MeV α particles and 120 MeV ~(12)C ions has been investigated. Maximum fluences were typically 10~(10)-10~(11) cm~(-2). Deconvolution techniques were applied to separate the glow curve into its component peaks. In all cases, the area of peak 3 showed a linear response as a function of fluence, for all the fluences studied while peaks 5 and 6 followed a linear-supralinear behaviour. The departure from linearity occurs at high doses, above 10~2 Gy. Concerning the dependence of the glow curve on type and energy of the incident radiation, the sensitivities of peaks 3 and 5 were studied as compared to the response to equal doses of gamma rays. The results, as a function of LET, follow the same general systematics observed previously by other authors. However, the quantitative comparison suggests that the linear energy transfer may not be the parameter which unifies all the observations.
机译:研究了CaF_2:Tm对0.7 MeV质子,5.3 MeVα粒子和120 MeV〜(12)C离子的TL响应。最大通量通常为10〜(10)-10〜(11)cm〜(-2)。应用了去卷积技术将辉光曲线分离为其分量峰值。在所有情况下,对于所有研究的通量,峰3的面积都显示出线性响应,它是注量的函数,而峰5和6遵循线性-超线性行为。在高剂量下(高于10〜2 Gy)会发生线性偏离。关于辉光曲线对入射辐射的类型和能量的依赖性,与对相同剂量的伽玛射线的响应相比,研究了峰3和峰5的灵敏度。结果,作为LET的函数,遵循其他作者先前观察到的相同的一般系统分析。但是,定量比较表明,线性能量传递可能不是统一所有观察结果的参数。