A PHANTOM-BASED EVALUATION OF THREE COMMERCIALLY AVAILABLE PATIENT ORGAN SHIELDS FOR COMPUTED TOMOGRAPHY X-RAY EXAMINATIONS IN DIAGNOSTIC RADIOLOGY

机译:用于诊断X线计算机断层摄影X射线检查的三种商业可用患者有机物的基于色度的评估

获取原文
获取原文并翻译 | 示例
           

摘要

Three commercially available in-plane patient organ shields (barium eye, bismuth eye and bismuth breast) for computed tomography (CT) examinations were evaluated to determine their effectiveness for dose reduction. Absorbed doses were measured using metal oxide semiconductor field effect transistor dosemeters fastened to a Kyoto CT Torso phantom. Resultant images were visually compared with those minus shielding by an experienced radiologist. Approximate dose reductions of 21, 38 and 50 % were achieved by the barium eye, bismuth eye and bismuth breast shields, respectively, at a cost of increased image noise and streak artefacts. Shielded images produced varied levels of image artefact, particularly those resulting from the eye shields. Measured dose reductions were not consistent with the potential dose savings stated by the manufacturers of the shields. When evaluating the breast shield, similar dose reduction was achieved without shield-induced artefact by simply reducing the X-ray tube current.
机译:对三种用于计算机断层扫描(CT)检查的市售面内患者器官防护罩(钡眼,铋眼和铋铋乳)进行了评估,以确定它们在降低剂量方面的有效性。使用固定在Kyoto CT Torso体模上的金属氧化物半导体场效应晶体管剂量计测量吸收剂量。经验丰富的放射线医师将所得图像与减负图像进行视觉比较。钡眼罩,铋眼罩和铋乳罩分别减少了约21%,38%和50%的剂量,但代价是增加了图像噪声和条纹假象。屏蔽的图像会产生各种级别的图像伪像,尤其是那些由眼罩产生的图像。测得的剂量减少量与防护罩制造商规定的可能的剂量节省量不一致。在评估乳房防护罩时,只需减少X射线管电流,就可以实现类似的剂量减少,而无需防护罩引起的假象。

著录项

  • 来源
    《Radiation Protection Dosimetry》 |2013年第2期|161-168|共8页
  • 作者单位

    Department of Clinical and Radiation Physics, Directorate of Healthcare Science and Technology, Plymouth Hospitals NHS Trust, Derriford Hospital, Plymouth, Devon PL6 8DH, UK;

    Imaging Directorate, Plymouth Hospitals NHS Trust, Derriford Hospital, Plymouth, Devon PL6 8DH, UK;

    Department of Clinical and Radiation Physics, Directorate of Healthcare Science and Technology, Plymouth Hospitals NHS Trust, Derriford Hospital, Plymouth, Devon PL6 8DH, UK;

  • 收录信息 美国《科学引文索引》(SCI);美国《化学文摘》(CA);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号