The initial response pattern for a conventional two-electrode, membrane-type dissolved-oxygen probe connected to an ENDECO model 1125 dissolved-oxygen sensor controller is presented. The controller implements a chronoamperometric technique and involves application of the polarizing potential as a pulse. The pulse duration confines the diffusion boundary within the internal electrolyte and the membrane. The time required for readout stabilization is attributed to the establishment of an equilibrium involving diffusion of reactant and product of the electrochemical reaction. A procedure for minimizing the role of membrane stretch in establishing the equilibrium time is discussed. Data obtained on changing the pulse repetition interval are presented in support of the establishment of an equilibrium condition. Sensor response over a 32-day period is noted. Calibration curve data demonstrate the linearity of response with varying oxygen concentrations. Response time data indicate attainment of better than 99% of full deflection within three minutes of a change in oxygen concentration.
机译:提出了连接到ENDECO 1125型溶解氧传感器控制器的常规两电极膜型溶解氧探头的初始响应模式。控制器实施计时电流技术,并涉及将极化电位作为脉冲施加。脉冲持续时间限制了内部电解质和膜内的扩散边界。读出稳定所需的时间归因于建立包括反应物扩散和电化学反应产物的平衡。讨论了最小化膜拉伸在建立平衡时间中的作用的方法。提供了有关改变脉冲重复间隔的数据,以支持建立平衡条件。记录了32天的传感器响应。校准曲线数据表明了氧气浓度变化时响应的线性。响应时间数据表明,在氧气浓度变化三分钟之内,可获得的挠度大于99%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号