Effects of the pre-irradiation storage procedure on the dose response of a Fricke xylenol orange gel dosimeter

机译:辐照前储存程序对Fricke二甲苯酚橙凝胶剂量计剂量响应的影响

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

The Fricke xylenol orange (FX) gel system is a chemical dosimeter characterized by good sensitivity, linear dose response, tissue equivalence, no toxicity, easy preparation, reproducibility and low cost. Thanks to the presence of the gelatinous matrix, the system is particularly suitable to perform reliable 3D mapping of the absorbed dose spatial distribution via magnetic resonance imaging (MRI) or optical techniques. The aim of this work is to study in a systematic way the influence of the pre-irradiation storage procedure upon sensitivity, dose response stability and lifetime of use of a FX gel system made with gelatin from porcine skin subjected to homogeneous irradiation. For this purpose, different pre-irradiation storage procedures, in terms of temperature and duration of each storage step, were investigated. In order to evaluate the dose response stability, the optical analyses of the samples were performed up to 6 hours after irradiation. Moreover, the samples were irradiated at time intervals of 24 hours for up to 7 days after preparation in order to evaluate the system lifetime of use. Regardless of their thermal and temporal life, the samples show linear dose responses in the investigated dose range (3-24 Gy) and an increase of sensitivity with the time elapsed between preparation and irradiation. Among the three pre-irradiation storage procedures considered here, a procedure that provides the best dose response stability and lifetime of use was identified and recommended for further use. The analyzed dosimetric system possesses good properties that make it promising for medical application, particularly concerning the evaluation of pre-treatment plan quality assurance within the conformational external beam radiotherapy.
机译:Fricke二甲苯酚橙(FX)凝胶系统是一种化学剂量计,具有良好的灵敏度,线性剂量响应,组织等效性,无毒性,易于制备,可重复性和低成本的特点。由于存在凝胶状基质,该系统特别适合通过磁共振成像(MRI)或光学技术对吸收的剂量空间分布进行可靠的3D映射。这项工作的目的是以系统的方式研究辐照前储存程序对使用明胶制成的FX凝胶系统的敏感性,剂量响应稳定性以及使用均一辐照的猪皮肤制成的FX凝胶系统的寿命的影响。为此,研究了在每个存储步骤的温度和持续时间方面不同的辐照前存储程序。为了评估剂量响应的稳定性,样品的光学分析在照射后长达6小时进行。此外,样品在制备后以24小时的时间间隔进行辐照,最多7天,以评估系统的使用寿命。无论其热寿命和时间寿命如何,样品在所研究的剂量范围(3-24 Gy)中均显示线性剂量响应,并且随着制备和照射之间时间的流逝,灵敏度增加。在此处考虑的三个辐照前存储程序中,确定了可提供最佳剂量响应稳定性和使用寿命的程序,并建议进一步使用。所分析的剂量系统具有良好的性能,使其在医学上具有广阔的应用前景,特别是在构象外照射治疗中评估治疗计划质量保证方面。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号