The dependencies of the light output response of CsI(Tl) crystals for various charged particle beams are investigated. Measurements were performed using 5.5 MeV ~(241)Am alpha particles, 220 MeV alpha and 110 MeV deuteron beams from the K500 Cyclotron at Texas A&M University, and p, ~2H, ~3He, ~6Li, and ~7Be beams from the Coupled Cyclotron Facility at Michigan State University. These measurements reveal a clear correlation between the non-uniformities observed in the same crystal for the various beams. Particle dependent global corrections were applied, resulting in final light output uniformities of the order of 0.1%. Annealed and non-annealed CsI(Tl) detectors and crystals with different thalium dopings are compared to determine whether annealing reduces the local variations in the light output. No significant correlation with the crystal fabrication process is observed.
机译:研究了CsI(Tl)晶体对于各种带电粒子束的光输出响应的依赖性。使用得克萨斯州A&M大学的K500回旋加速器的5.5 MeV〜(241)Am alpha粒子,220 MeV alpha和110 MeV氘核束以及来自耦合的p,〜2H,〜3He,〜6Li和〜7Be束进行测量密歇根州立大学的回旋加速器。这些测量结果揭示了在同一晶体中观察到的各种光束之间的不均匀性之间存在明显的相关性。应用了依赖于粒子的全局校正,导致最终的光输出均匀性约为0.1%。比较退火和未退火的CsI(Tl)检测器以及具有不同掺杂的晶体,以确定退火是否减少了光输出的局部变化。没有观察到与晶体制造过程的显着相关性。