首页> 外文期刊>Nuclear Instruments & Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment >Monte Carlo and least-squares methods applied in unfolding of X-ray spectra measured with cadmium telluride detectors
【24h】

Monte Carlo and least-squares methods applied in unfolding of X-ray spectra measured with cadmium telluride detectors

机译:蒙特卡洛和最小二乘方法用于碲化镉探测器测量的X射线光谱的展开

获取原文
获取原文并翻译 | 示例
           

摘要

Spectra of calibration sources and X-ray beams were measured with a cadmium telluride (CdTe) detector. The response function of the detector was simulated using the GEANT4 Monte Carlo toolkit. Trapping of charge carriers were taken into account using the Hecht equation in the active zone of the CdTe crystal associated with a continuous function to produce drop of charge collection efficiency near the metallic contacts and borders. The rise time discrimination is approximated by a cut in the depth of the interaction relative to cathode and corrections that depend on the pulse amplitude. The least-squares method with truncation was employed to unfold X-ray spectra typically used in medical diagnostics and the results were compared with reference data.
机译:使用碲化镉(CdTe)检测器测量校准源和X射线束的光谱。使用GEANT4蒙特卡洛工具包模拟了检测器的响应功能。使用Hecht方程考虑了CdTe晶体活性区域中与连续函数相关的电荷载流子的捕获,从而在金属触点和边界附近产生电荷收集效率的下降。通过减少相对于阴极的相互作用深度并根据脉冲幅度进行校正,可以近似得出上升时间的区别。采用截短的最小二乘法来展开通常用于医学诊断的X射线光谱,并将结果与​​参考数据进行比较。