首页> 外文期刊>Nuclear instruments and methods in physics research >Synthetic Single Crystal Diamond Diodes For Radiotherapy Dosimetry
【24h】

Synthetic Single Crystal Diamond Diodes For Radiotherapy Dosimetry

机译:放射剂量法合成单晶金刚石二极管

获取原文
获取原文并翻译 | 示例
           

摘要

Synthetic single crystal diamonds in a p-type/intrinsic/metal structure were tested as dosimeters for radiotherapy. The devices have been analyzed by using 6 and 10 MV Bremsstrahlung X-ray beams and electron beams from 6 MeV up to 18 MeV from a CUNAC DHX Varian accelerator. All measurements have been performed in a water phantom and ionization chambers were used for calibration and comparison. The dosimeters were operated in photovoltaic regime with no external bias voltage applied. A few Gy pre-irradiation was performed in order to stabilize the device output, resulting in fluctuations sensitivity below ±0.5%. No dose rate dependence of the detector response was observed. Very good reproducibility and linearity were obtained as well.
机译:测试了p型/本征/金属结构的合成单晶金刚石作为放射疗法的剂量计。通过使用CUNAC DHX Varian加速器从6 MeV到18 MeV的6和10 MV ms致辐射X射线束和电子束对设备进行了分析。所有测量均在水模中进行,电离室用于校准和比较。剂量计在未施加外部偏置电压的光伏条件下运行。为了稳定设备输出,进行了几次Gy预照射,导致波动灵敏度低于±0.5%。没有观察到检测器响应的剂量率依赖性。还获得了非常好的重现性和线性。