Dry Dilution Refrigerators (DDR) based on pulse tube cryo-coolers have started to replace Wet Dilution Refrigerators (WDR) due to the ease and low cost of operation. However these advantages come at the cost of increased vibrations, induced by the pulse tube. In this work, we present the vibration measurements performed on three different commercial DDRs. We describe in detail the vibration measurement system we assembled, based on commercial accelerometers, conditioner and DAQ, and examined the effects of the various damping solutions utilized on three different DDRs, both in the low and high frequency regions. Finally, we ran low temperature, pseudo-massive (30 and 250 g) germanium bolometers in the best vibration-performing system under study and report on the results.
机译:基于脉冲管低温冷却器的干式稀释冰箱(DDR)由于操作简便且成本低廉,已开始取代湿式稀释冰箱(WDR)。但是,这些优点是以脉冲管引起的振动增加为代价的。在这项工作中,我们介绍了在三种不同的商用DDR上执行的振动测量。我们详细描述了基于商业加速度计,调节器和DAQ组装的振动测量系统,并研究了在低频和高频区域对三种不同DDR使用的各种阻尼解决方案的影响。最后,我们在正在研究的最佳振动执行系统中运行了低温准假体(30 g和250 g)锗测辐射热仪,并报告了结果。