首页> 外文期刊>Microchimica Acta >Automatic Fluorimetric Procedure for the Determination of Aluminium in Plant Nutrient Solution and Natural Water Employing a Multicommutated Flow System
【24h】

Automatic Fluorimetric Procedure for the Determination of Aluminium in Plant Nutrient Solution and Natural Water Employing a Multicommutated Flow System

机译:采用多换向流动系统的自动荧光法测定植物营养液和天然水中的铝

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

A fluorimetric flow procedure for the determination of aluminium in plant nutrient solutions and natural waters using morin (3,5,7,2′,4′-(pentahydroxyflavone) as a fluorimetric reagent is described. The flow system comprised a set of three-way solenoid valves that were assembled to implement the multicommutation and monosegmented flow approaches. The calibration graph was linear in the range between 0.1 to 1.0 mg L−1 Al3+. A relative standard deviation of 1.7% (n=7) for a typical sample containing 0.5 mg L−1 Al was achieved. Comparing the results with those obtained by atomic emission spectrometry (ICP-AES), no significant difference at the 95% confidence level was observed. Other profitable features such as an analytical throughput of 60 determinations per hour; a reagent consumption of 1.6 µg of morin per determination and a detection limit of 0.04 mg L−1 were also achieved.
机译:介绍了一种荧光流动法,用于测定植物营养液和天然水中的铝,其中使用了rin(3,5,7,2',4'-(五羟基黄酮)作为荧光试剂。组装以实现多换向和单段流量方法的单向电磁阀。校准曲线在0.1至1.0 mg L-1 Al3 + 范围内呈线性关系,相对标准偏差为1.7%(n = 7)对于典型的含0.5 mg L-1 Al的样品,获得的结果与原子发射光谱法(ICP-AES)的结果相比,在95%置信度水平上没有观察到显着差异。有利的功能,如每小时分析60次,分析通量,每次测定消耗的试剂量为1.6 µg morin和检测限为0.04 mg L-1