Tooth Form Evaluation Using Ball Artifact Development of a Measuring Instrument of a Ball Center Distance Traceable to National Standard of Length

机译:利用可追溯至国家标准长度的球心距测量仪的球工件开发齿形评估

获取原文
获取原文并翻译 | 示例
           

摘要

AIST (National Institute of Advanced Industrial Science and Technology) has proposed a new calibration method of gear measuring instrument using a ball artifact consisting of two master balls and an optical flat base (DBA). We measure a circular arc instead of involute tooth form. Difference of the distance between the arc of the master ball and involute curve can be calculated theoretically. For the calculation, center distance between two balls is required and the value of measurement uncertainty is less than 0.1 μm and also should be traceable to national standard of length. This paper reports on a new method and an originally developed instrument to measure the center distance between two balls using laser interferometer. The expanded uncertainty in measurement of a center distance of 44 mm was 21 nm.
机译:AIST(国家先进工业科学技术研究院)提出了一种齿轮测量仪器的新校准方法,该方法使用了由两个主球和一个光学平底(DBA)组成的球假象。我们测量的是圆弧形而不是渐开线齿形。理论上可以计算出主球弧线与渐开线曲线之间的距离之差。为了进行计算,需要两个球之间的中心距离,并且测量不确定度的值小于0.1μm,并且还可以追溯到国家标准的长度。本文介绍了一种使用激光干涉仪测量两个球之间的中心距离的新方法和一种原始开发的仪器。中心距44 mm的测量中扩展的不确定度为21 nm。