Purpose: This study evaluated the difference in verification performance and organ doses between megavoltage computed tomography (MVCT) and kilovoltage cone-beam computed tomography (kV CBCT).Methods: Anthropomorphic phantoms of head-and-neck (H&N) and pelvic regions were scanned by the computed tomography–simulator. A common reference standard setup position for each phantom was determined. Markings of known deviations from the standard position in the lateral, longitudinal and yaw displacements were made on the phantoms. The verifications by kV CBCT and MVCT were conducted in the linear accelerator and helical tomotherapy treatment unit respectively. The phantoms were then shifted in steps according to the assigned different degree of positional deviations. The discrepancy between the detected data and the known data (δD) indicated the detection error of the verification system. Thermoluminescent dosimeters were used for dose measurements of kV CBCT and MVCT.Results: δD of the lateral translations and yaw rotation in kV CBCT of the H&N region were lower than MVCT. There were no differences in the error detection along longitudinal direction between the two verification methods. For H&N, the mean doses of the various organs were significantly lower in kV CBCT except the skin. For the pelvis, MVCT delivered significantly higher mean dose to the prostate and femoral heads.
机译:目的:这项研究评估了兆伏计算机断层扫描(MVCT)和千伏锥形束计算机断层扫描(kV CBCT)在验证性能和器官剂量方面的差异。通过计算机断层扫描模拟器。确定了每个体模的通用参考标准设置位置。在人体模型上标记了在横向,纵向和偏航位移中偏离标准位置的已知偏差。通过kV CBCT和MVCT进行的验证分别在线性加速器和螺旋层析疗法治疗单元中进行。然后根据分配的不同程度的位置偏差逐步移动幻影。检测到的数据与已知数据之间的差异(δD)表明验证系统的检测错误。结果表明:H&N区kV CBCT的横向平移和偏航旋转的δD均低于MVCT。两种验证方法之间沿纵向方向的错误检测没有差异。对于H&N,除皮肤外,kV CBCT中各个器官的平均剂量均显着降低。对于骨盆,MVCT向前列腺和股骨头输送的平均剂量明显较高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号