首页> 外文期刊>Journal of Material Cycles and Waste Managemen >Environmental Technology Verification (ETV) test of dioxin emission monitors
【24h】

Environmental Technology Verification (ETV) test of dioxin emission monitors

机译:二恶英排放监测仪的环境技术验证(ETV)测试

获取原文
获取原文并翻译 | 示例
           

摘要

The performance of four dioxin emission monitors, including two long-term sampling devices, the Dioxin-MonitoringSystem (DMS) and AMESA (the adsorption method for sampling dioxins and furans), and two semi-real-time continuous monitors, the resonance ionization with multimirror photon accumulation time-of-flight mass spectrometer (RIMMPA-TOFMS) and the jet resonance-enhanced multiphoton ionization (jet-REMPI) system were tested. A package boiler burning a simulated chlorinated hazardous waste was used for a total of nine tests. Reference samples were collected during each test and analyzed for polychlorinated dibenzodioxins and dibenzofurans (PCDDs/Fs) using gas chromatography mass spectrometry. The PCDD/F concentrations of the reference samples measured by EPA Method 23 ranged from 0.9 to 60ng toxic equivalence (TEQ)/dry standard cubic meter. The relative accuracies achieved by DMS, AMESA, and jet-REMPI varied from 22.6% to 78.2%, with 100% data completeness. The RIMMPA-TOFMS produced no quantifiable results due to various difficulties associated with the instrument during the testing. The two long-term samplers were easy to install and operate and provided a cumulative, averaged emission for the sampling period. The operations of the two semi-real-time continuous monitors were relatively complex, but one of them provided on-site, real-time data for PCDD/ F emissions from measurement of a TEQ correlative indicator compound. This article summarizes results from the individual Environmental Technology Verification reports for the four dioxin monitors.
机译:四个二恶英排放监测仪的性能,包括两个长期采样设备,即二恶英监测系统(DMS)和AMESA(用于采样二恶英和呋喃的吸附方法),以及两个半实时连续监测仪,通过测试了多镜光子累积飞行时间质谱仪(RIMMPA-TOFMS)和喷射共振增强多光子电离(jet-REMPI)系统。燃烧模拟氯化危险废物的成套锅炉共进行了九项测试。在每个测试期间收集参考样品,并使用气相色谱质谱法分析多氯联苯二恶英和二苯并呋喃(PCDDs / Fs)。通过EPA方法23测量的参考样品的PCDD / F浓度为0.9至60ng毒性当量(TEQ)/干标准立方米。 DMS,AMESA和jet-REMPI的相对精度从22.6%到78.2%不等,数据完整性为100%。由于在测试过程中与仪器相关的各种困难,RIMMPA-TOFMS无法产生可量化的结果。这两个长期采样器易于安装和操作,并在采样期间提供了累积的平均排放量。两个半实时连续监测器的操作相对复杂,但是其中之一提供了通过测量TEQ相关指示剂化合物获得的PCDD / F排放的实时数据。本文总结了四个二恶英监测仪各自的环境技术验证报告的结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号