首页> 外文期刊>Journal of Hazardous Materials >Determination of nanomolar chromate in drinking water with solid phase extraction and a portable spectrophotometer
【24h】

Determination of nanomolar chromate in drinking water with solid phase extraction and a portable spectrophotometer

机译:固相萃取-便携式分光光度计测定饮用水中的纳摩尔铬酸盐

获取原文
获取原文并翻译 | 示例
       

摘要

Determination of chromate at low concentration levels in drinking water is an important analytical objective for both human health and environmental science. Here we report the use of solid phase extraction (SPE) in combination with a custom-made portable light-emitting diode (LED) spectrophotometer to achieve detection of chromate in the field at nanomolar levels. The measurement chemistry is based on a highly selective reaction between 1,5-diphenylcarbazide (DPC) and chromate under acidic conditions. The Cr-DPC complex formed in the reaction can be extracted on a commercial C18 SPE cartridge. Concentrated Cr-DPC is subsequently eluted with methanol and detected by spectrophotometry. Optimization of analytical conditions involved investigation of reagent compositions and concentrations, eluent type, flow rate (sample loading), sample volume, and stability of the SPE cartridge. Under optimized conditions, detection limits are on the order of 3 nM. Only 50 mL of sample is required for an analysis, and total analysis time is around l0min. The targeted analytical range of 0-500 nM can be easily extended by changing the sample volume. Compared to previous SPE-based spectrophotometric methods, this analytical procedure offers the benefits of improved sensitivity, reduced sample consumption, shorter analysis time, greater operational convenience, and lower cost.
机译:饮用水中低浓度铬酸盐的测定是人类健康和环境科学的重要分析目标。在这里,我们报告了固相萃取(SPE)与定制的便携式发光二极管(LED)分光光度计结合使用以实现纳摩尔浓度水平的现场铬酸盐的检测。测量化学基于1,5-二苯卡巴肼(DPC)与铬酸盐在酸性条件下的高度选择性反应。反应中形成的Cr-DPC络合物可以在商用C18 SPE柱上提取。随后用甲醇洗脱浓缩的Cr-DPC,并通过分光光度法进行检测。分析条件的优化涉及对试剂组成和浓度,洗脱液类型,流速(样品上样),样品量和SPE柱的稳定性的研究。在优化条件下,检测极限约为3 nM。分析仅需要50 mL样品,总分析时间约为10分钟。通过更改样品量可以轻松扩展0-500 nM的目标分析范围。与以前的基于SPE的分光光度法相比,该分析程序具有以下优势:灵敏度提高,样品消耗减少,分析时间更短,操作方便,成本更低。

著录项

  • 来源
    《Journal of Hazardous Materials》 |2012年第15期|p.247-252|共6页
  • 作者单位

    College of Marine Science, University of South Florida, 140 7th Avenue South, St. Petersburg, FL 33701, United States;

    College of Marine Science, University of South Florida, 140 7th Avenue South, St. Petersburg, FL 33701, United States;

    College of Marine Science, University of South Florida, 140 7th Avenue South, St. Petersburg, FL 33701, United States;

  • 收录信息 美国《科学引文索引》(SCI);美国《工程索引》(EI);美国《生物学医学文摘》(MEDLINE);美国《化学文摘》(CA);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    solid phase extraction; nanomolar detection limit; chromate; carbon 18 cartridge; light-emitting diode;

    机译:固相萃取纳摩尔检测限铬酸盐;碳18墨盒;发光二极管;
  • 入库时间 2022-08-17 13:23:10