We present the application of nanosecond laser ablation (LA) coupled to a 'Nu Plasma HR' multi collector inductively coupled plasma mass spectrometer (MC-ICP-MS) for the direct analysis of U isotope ratios in single, 10-20 μm-sized, U-doped glass particles. Method development included studies with respect to (1) external correction of the measured U isotope ratios in glass particles, (2) the applied laser ablation carrier gas (i.e. Ar versus He) and (3) the accurate determination of lower abundant ~(236)U/~(238)U isotope ratios (i.e. 10~(-5)). In addition, a data processing procedure was developed for evaluation of transient signals, which is of potential use for routine application of the developed method. We demonstrate that the developed method is reliable and well suited for determining U isotope ratios of individual particles. Analyses of twenty-eight S1 glass particles, measured under optimized conditions, yielded average biases of less than 0.6% from the certified values for ~(234)U/~(238)U and 235U/~(238)U ratios. Experimental results obtained for ~(236)U/~(238)U isotope ratios deviated by less than -2.5% from the certified values. Expanded relative total combined standard uncertainties U_c(k = 2) of 2.6%, 1.4% and 5.8% were calculated for ~(234)U/~(238)U, ~(235)U/~(238)U and ~(236)U/~(238)U, respectively.
机译:我们介绍了纳秒激光烧蚀(LA)与'Nu Plasma HR'多收集器电感耦合等离子体质谱仪(MC-ICP-MS)结合的应用,用于直接分析10-20μm大小的U同位素比率,掺U的玻璃颗粒。方法开发包括以下方面的研究:(1)对玻璃颗粒中测得的U同位素比进行外部校正;(2)所施加的激光烧蚀载气(即Ar与He的关系);以及(3)准确测定较低的丰度〜(236) U /〜(238)U同位素比(即10〜(-5))。另外,开发了用于评估瞬态信号的数据处理程序,这对于所开发方法的常规应用可能具有潜在用途。我们证明开发的方法是可靠的,并且非常适合确定单个粒子的U同位素比。在最佳条件下测量的28个S1玻璃粒子的分析得出,相对于〜(234)U /〜(238)U和235U /〜(238)U比的认证值,平均偏差小于0.6%。 〜(236)U /〜(238)U同位素比与标准值相差小于-2.5%的实验结果。计算〜(234)U /〜(238)U,〜(235)U /〜(238)U和〜()的扩展相对总合并标准不确定度U_c(k = 2)分别为2.6%,1.4%和5.8%。 236)U /〜(238)U。

著录项

  • 来源
    《Journal of Environmental Radioactivity》 |2012年第2012期|8-15|共8页
  • 作者单位

    University of Natural Resources and Life Sciences Vienna, Department of Chemistry, Division of Analytical Chemistry, VIRIS-Laboratory, Konrad-Lorenz-Strasse 24, 3430 Tulln, Austria;

    Safeguards Analytical Services, Department of Safeguards, International Atomic Energy Agency, Wagramer Strasse 5, 1400 Vienna, Austria;

    University of Natural Resources and Life Sciences Vienna, Department of Chemistry, Division of Analytical Chemistry, VIRIS-Laboratory, Konrad-Lorenz-Strasse 24, 3430 Tulln, Austria;

  • 收录信息 美国《科学引文索引》(SCI);美国《工程索引》(EI);美国《生物学医学文摘》(MEDLINE);美国《化学文摘》(CA);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Laser ablation; Multi collector ICP-MS; Uranium isotope ratio analysis; Particle analysis; Nuclear forensics; Nuclear safeguards;

    机译:激光烧蚀;多收集器ICP-MS铀同位素比分析;颗粒分析;核法证学;核保障;