首页> 外文期刊>Journal of Analytical Atomic Spectrometry >Electrospray deposition followed by laser-induced breakdown spectroscopy (ESD-LIBS): a new method for trace elemental analysis of aqueous samples
【24h】

Electrospray deposition followed by laser-induced breakdown spectroscopy (ESD-LIBS): a new method for trace elemental analysis of aqueous samples

机译:电喷雾沉积,然后进行激光诱导击穿光谱分析(ESD-LIBS):一种用于水样中痕量元素分析的新方法

获取原文
获取原文并翻译 | 示例
           

摘要

The combination of electrospray deposition with laser-induced breakdown spectroscopy (ESD-LIBS) was investigated as a potential approach to the detection and quantification of Zn, Cd, Cr and Ni, at trace levels, in aqueous samples. In this analytical procedure, the aqueous samples were first converted into solids in an electrospray system. To this end, micro-volumes of liquids were electro-sprayed onto a heated substrate, leading to the generation of solid residues. Afterwards, the as-obtained residues were analysed by LIBS. Three calibration methodologies were tested with the proposed ESD-LIBS methodology, namely external calibration, conventional standard addition calibration and on-line standard addition calibration. In all cases, the analytical features of the ESD-LIBS method were assessed. The obtained limits of detection ranged from 9 mu g kg(-1) to 57 mu g kg(-1), depending on the element and on the calibration modality used. Method trueness, evaluated from the analysis of a real sample of tap water, was highly dependent on the calibration method. The use of external calibration led to recovery values in the range of 18-123%, indicating the existence of strong matrix effects. This drawback was overcome with the application of conventional standard addition and on-line standard addition calibration modalities, for which recovery values were improved to the ranges 91-110% and 90-105%, respectively. Among them, the use of on-line standard addition provides a sensitive and accurate methodology with possibilities of automation.
机译:研究了电喷雾沉积与激光诱导击穿光谱法(ESD-LIBS)的结合,将其作为检测和定量痕量水样中Zn,Cd,Cr和Ni的潜在方法。在该分析程序中,首先将水样品在电喷雾系统中转化为固体。为此,将微量的液体电喷雾到加热的基材上,导致固体残留物的产生。然后,通过LIBS分析所获得的残留物。使用提议的ESD-LIBS方法测试了三种校准方法,即外部校准,常规标准添加校准和在线标准添加校准。在所有情况下,都评估了ESD-LIBS方法的分析功能。所获得的检出限在9μg kg(-1)到57μg kg(-1)之间,具体取决于元素和所使用的校准方式。通过对自来水真实样品的分析评估得出的方法真实性高度依赖于校准方法。外部校准的使用导致回收率在18-123%的范围内,表明存在强基质效应。通过应用常规标准添加和在线标准添加校准方法克服了这一缺点,将其回收率分别提高到91-110%和90-105%。其中,使用在线标准添加提供了灵敏而准确的方法,并具有自动化的可能性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号