首页> 外文期刊>Journal of Analytical Atomic Spectrometry >On-line preconcentration and determination of vanadium in tap and river water samples by flow injection-inductively coupled plasma-optical emission spectrometry (FI-ICP-OES)
【24h】

On-line preconcentration and determination of vanadium in tap and river water samples by flow injection-inductively coupled plasma-optical emission spectrometry (FI-ICP-OES)

机译:在线预富集和流动注射-电感耦合等离子体发射光谱法(FI-ICP-OES)测定自来水和河水中的钒

获取原文
获取原文并翻译 | 示例
           

摘要

A flow injection system coupled to an inductively coupled plasma optical emission spectrometer (ICP-OES) was used for the on-line preconcentration and subsequent determination of vanadium. Trace amounts of V were preconcentrated by sorption on a conical minicolumn packed with immobilised yeast cells in the absence of complexing reagent at pH 7.0. Vanadium was removed from the minicolumn with a 50% HCl solution. An enrichment factor of 45-fold for a sample volume of 10 ml was obtained. The detection limit (LoD) for the preconcentration method proposed was 0.06 mu g l(-1). The precision for 10 replicate determinations at the 5 mu g l(-1) V level was 3.4% relative standard deviation (RSD), calculated from the peak heights obtained. The calibration graph of the preconcentration method for V was linear with a correlation coefficient of 0.9997 at levels near the detection limits up to at least 100 mu g l(-1). The method was successfully applied to the determination of V in tap and river water samples.
机译:与电感耦合等离子体发射光谱仪(ICP-OES)耦合的流动注射系统用于在线预富集和随后测定钒。在pH 7.0下不存在络合剂的情况下,通过吸附在装有固定化酵母细胞的圆锥形微型柱上,对痕量的V进行预浓缩。用50%HCl溶液从小柱中除去钒。对于10 ml的样品量,获得了45倍的富集因子。提出的预浓缩方法的检测限(LoD)为0.06μg l(-1)。根据获得的峰高计算,在5μg l(-1)V浓度下进行10次重复测定的精度为3.4%相对标准偏差(RSD)。 V的预富集方法的校正图呈线性,相关系数为0.9997,在接近检出限的水平上至少为100μg l(-1)。该方法已成功地用于自来水和河水中样品中V的测定。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号