Measurements of Dose-Averaged Linear Energy Transfer Distributions in Water Using CR-39 Plastic Nuclear Track Detector for Therapeutic Carbon Ion Beams

机译:使用治疗性碳离子束的CR-39塑料核径迹检测器测量水中的剂量平均线性能量转移分布

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A CR-39 plastic nuclear track detector was used as a linear energy transfer (LET) detector for carbon ion radiotherapy. We compared dose-averaged LET distributions in water obtained using the CR-39 detector for a monoenergetic beam and spread-out Bragg peak beam by calculations using the one-dimensional heavy-ion transport code used in the current heavy-ion treatment planning. We confirmed that the CR-39 detector could measure the high LET particles that are dominant contributors to dose-averaged LET. On the other hand, the CR-39 result was overestimated in the tail region of the distal edge in depth-dose distributions, due to its detection limit for lower LET particles. However, physical dose in the region is quite small. Namely, the effect of this difference on the biological dose distribution is also small. These results demonstrate that the CR-39 detector is a useful detector for measuring the LET distribution in carbon ion radiotherapy.
机译:CR-39塑料核径迹检测器用作碳离子放射疗法的线性能量转移(LET)检测器。通过使用当前重离子处理计划中使用的一维重离子迁移代码进行计算,我们比较了使用CR-39检测器获得的水中单剂量束和分散布拉格峰束的剂量平均LET分布。我们证实,CR-39检测器可以测量高LET粒子,这些粒子是剂量平均LET的主要贡献者。另一方面,由于其对较低的LET粒子的检测极限,CR-39结果在远侧边缘的尾部区域的深度剂量分布被高估了。但是,该区域的物理剂量很小。即,这种差异对生物剂量分布的影响也很小。这些结果表明,CR-39检测器是用于测量碳离子放射治疗中LET分布的有用检测器。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号