首页> 外文期刊>International Orthopaedics >Deformation of metal-backed acetabular components and the impact of liner thickness in a cadaveric model
【24h】

Deformation of metal-backed acetabular components and the impact of liner thickness in a cadaveric model

机译:尸体模型中金属支撑髋臼组件的变形和衬里厚度的影响

获取原文
获取原文并翻译 | 示例
       

摘要

Shell deformation of resurfacing and all-metal modular cups following press-fit implantation has been reported, but not for conventional metal-backed cups with polyethylene liners. The deformation of acetabular components with historical and thin polyethylene inserts after press-fit insertion was evaluated using a cadaveric model. All shells and liners deformed upon implantation. Following joint loading, shell pinch decreased from 0.32 to 0.22 mm (p = 0.019) and from 0.29 to 0.13 mm (p = 0.003) for the thin and thick liner groups, respectively. Liner pinch also decreased from 0.17 to 0.04 mm (p = 0.031) and from 0.06 to 0 mm (p = 0.103) for the thin and thick liner groups, respectively. There were no significant differences between the thin and thick liners. Liner deformation was influenced by the initial shell deformation and donor bone quality. Shell and liner pinch decreased following joint loading, suggesting a settling in effect.
机译:已经报道了压配合植入后重铺表面和全金属模块化杯的外壳变形,但是没有传统的带有聚乙烯衬里的金属背杯的变形。使用尸体模型评估了压入配合后具有历史和薄聚乙烯插入物的髋臼组件的变形。所有的外壳和衬里在植入时都会变形。接缝加载后,薄衬套和厚衬套组的壳夹缩分别从0.32毫米减小到0.22毫米(p = 0.019),从0.29减小到0.13毫米(p = 0.003)。薄衬套和厚衬套组的衬套收缩也分别从0.17毫米降低到0.04毫米(p = 0.031),从0.06降低到0毫米(p = 0.103)。薄衬里和厚衬里之间没有显着差异。内衬变形受初始外壳变形和供体骨质量的影响。接缝加载后,壳和衬套的收缩减少,表明沉降有效。