首页> 外文期刊>International Journal of Thermophysics >Calibration of the Irradiance Responsivity of a Filter Radiometer for T Measurement at NIM
【24h】

Calibration of the Irradiance Responsivity of a Filter Radiometer for T Measurement at NIM

机译:NIM中用于T测量的滤光片辐射计的辐照度响应的校准

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

At the National Institute of Metrology of China (NIM), silicon photodiode-based narrow-band interference filter radiometers (FRs) have been designed for the radiometric determination of the thermodynamic temperature. The FR calibrations were performed on a new spectral comparator with a trap detector which was calibrated against the cryogenic radiometer at several discrete laser lines. The new spectral comparator is constructed from two grating monochromators assembled to give lower stray light and higher transmitted flux. Applying a transmittance measurement of the filter in the out-of-band region and careful control of the temperature, the irradiance responsivity of a 633 nm centered FR has been obtained over a dynamic range of nearly eight decades in the wavelength range from 450 nm to 1200 nm. The relative standard uncertainty of the responsivity is also analyzed and estimated to be less than 7 × 10−4 at the 1 σ level.
机译:在中国国家计量学院(NIM),已经设计了基于硅光电二极管的窄带干涉滤光片辐射计(FR)来进行热力学温度的辐射测定。 FR校准是在带有陷波检测器的新光谱比较器上进行的,该陷波检测器是针对低温辐射计在几条离散激光线上进行校准的。新的光谱比较器是由两个光栅单色仪组成的,这些单色仪组装在一起以提供较低的杂散光和较高的透射通量。通过对带外区域的滤光片进行透射率测量并仔细控制温度,在从450 nm至450 nm的波长范围内的近八十年的动态范围内,获得了633 nm中心FR的辐照度1200海里还对响应度的相对标准不确定度进行了分析,并估计其在1σ水平下小于7×10 -4