Nanotrace Cd determination in natural waters using two new flotation collectors, lead(Ⅱ) heptyldithio-carbamate, Ph(HpDTC)_2, and cobalt(Ⅲ) heptyldithiocarbamate, Co(HpDTC)_3, are presented. The optimization of the most important experimental parameters for flotation is given. Zeeman electrothermal atomic absorption spectrometry (ZETAAS) is used as an instrumental technique for Cd measurement. The results are compared with those obtained by other preconcentration procedures as independent methods. The limit of detection of ZETAAS using Pb(HpDTC)_2 as collector is 4.8ng/L, while using Co(HpDTC)_3 it is 3.0ng/L.
机译:提出了使用两种新型浮选捕集剂铅(Ⅱ)庚基二硫代氨基甲酸铅(Ph(HpDTC)_2)和钴(Ⅲ)庚基二硫代氨基甲酸钴(Co(HpDTC)_3)进行纳米痕量镉的测定。给出了浮选最重要的实验参数的优化。塞曼电热原子吸收光谱法(ZETAAS)被用作镉测量的仪器技术。将结果与通过其他预浓缩程序作为独立方法获得的结果进行比较。使用Pb(HpDTC)_2作为收集剂的ZETAAS的检出限为4.8ng / L,而使用Co(HpDTC)_3的检出限为3.0ng / L。