首页> 外文期刊>International Journal of Computers & Applications >HAEMATOMA DETECTION USING EIT IN A SHEEP MODEL
【24h】

HAEMATOMA DETECTION USING EIT IN A SHEEP MODEL

机译:在绵羊模型中使用EIT进行血肿检测

获取原文
获取原文并翻译 | 示例
       

摘要

Performance evaluation of a portable digital electrical impedance tomography system to detect haematomas using phantom and sheep models is presented. Experiments have been performed using 8-electrode full array configuration. Conductivity changes were detected in phantom model while anomalies were placed at centre and close to the edge of the tank. Bleeding rate was successfully monitored in sheep model while blood-like conductivity solution was injected via the brainstem. EIT images were reconstructed sequentially for different injection volumes and the quantity index (QI) was calculated as a function of the injected solution volume. The results show a linear relationship of QI to the injected volume. Images of the sheep experiment with the simulated haematomas, blood-like conductivity gel, placed on top of the parietal lobes of the brain on the left and right sides were reconstructed and haematomas' are clearly detected and localized. These experiments prove that the detection and quantification of haematomas in the brain is possible and encourage further investigation for medical applications.
机译:提出了使用幻影和绵羊模型的便携式数字电阻抗层析成像系统检测血肿的性能评估。已经使用8电极全阵列配置进行了实验。在将幻影放置在水箱的中心和边缘附近时,在幻影模型中检测到电导率变化。通过脑干注入血液样电导率溶液后,成功监测了绵羊模型的出血率。针对不同的进样量依次重建EIT图像,并根据进样溶液量计算数量指数(QI)。结果表明QI与注入量之间存在线性关系。用模拟的血肿,血液样电导率凝胶放置在左侧和右侧大脑顶叶顶部的绵羊实验图像被重建,血肿被清晰地检测和定位。这些实验证明,脑中血肿的检测和定量是可能的,并鼓励进一步研究医学应用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号